FDA Medical Device Recalls (openFDA)
Zimmer Inc.: Nextgen Complete Knee Solution Replacement Jaw for Use with Femoral and Provisional Impactor/Extractor, REF 00-5901-026-40, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components.
- Recall number
- Z-2092-2012
- Recalling firm
- Zimmer Inc.
- Product
- Nextgen Complete Knee Solution Replacement Jaw for Use with Femoral and Provisional Impactor/Extractor, REF 00-5901-026-40, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components.
- Status
- Terminated
- Initiated
- 2012-04-30
- Posted
- 2012-07-27
- Terminated
- 2014-06-04
- Reason
- The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall
- Root cause
- Component design/selection
- Action
- Zimmer sent an Urgent Medical Device Recall letter dated April 30, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Requests for alternative instruments should be requested by calling (574) 372-4470. Fax Back Inventory Return Certification Form is to be sent back to Zimmer at (574) 372-4265. All recalled instruments available at the distributor are to be returned to Zimmer Product Service Department, 1777 West Center Street, Warsaw, IN 46580. The Distributor notice included instructions to notify surgeons immediately of the recall and requested a list of hospitals and surgeons receiving the device be sent back to Zimmer via E-mail. Zimmer is continuing distribution of the replacement jaw ( 005901-026-40) until alternative tools are available. The distributors are asked to identify and deliver preferred alternative instruments to affected accounts. Beginning May 7, 2012 Urgent Recall notices are being sent to hospitals and surgeons identified with instructions to acknowledge the recall communication. For questions regarding this recall call 574-372-4807.
- Quantity
- 65 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Singapore, China, Hong Kong, Canada, Australia, Japan, Europe, Germany, India, and Finland.
- Lot, serial or model codes
- Part 00-5901-026-40 Lot 62011774, 61870067, and 61813699 Initially the 00-5901-026-40 replacement jaws will remain available for order until all of the alternative instruments are in place, however they will also be required to be returned. The final lot numbers and manufacture dates will be provided in a subsequent status update.
- 510(k) numbers
- ["K031061"]
- PMA numbers
- not on file
- Product code
- MBH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28