FDA Medical Device Recalls (openFDA)
Synthes (USA) Products LLC: Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element insertion. Manual orthopedic surgical instrument.
- Recall number
- Z-2091-2015
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element insertion. Manual orthopedic surgical instrument.
- Status
- Terminated
- Initiated
- 2015-06-10
- Posted
- 2015-07-16
- Terminated
- 2016-08-18
- Reason
- The height of one of the three guiding pins of the Helical Blade Inserter for the TFNA is oversized. The Helical Blade Inserter is therefore unable to pass through the Blade/Screw Guide Sleeve (part 03.037.017).
- Root cause
- Process control
- Action
- A recall notification letter, dated June 10, 2015, was sent to sales consultants and Synthes distribution sites requesting return of affected devices.
- Quantity
- 32
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Part Number Lot Numbers 03.037.024 T102759
- 510(k) numbers
- ["K131548"]
- PMA numbers
- not on file
- Product code
- HSB
- Firm FEI
- 3008812560
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06