Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Syngo.plaza, Picture archiving and communication system (PACS), Model Numbers - 10863171, 10863172, 10863173

Recall number
Z-2088-2016
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Syngo.plaza, Picture archiving and communication system (PACS), Model Numbers - 10863171, 10863172, 10863173,
Status
Terminated
Initiated
2016-05-09
Posted
2016-06-29
Terminated
2017-03-20
Reason
Software error in previous software versions in which two references for the same image may exist in the database.
Root cause
Software design
Action
Siemens sent an Customer Information letter dated May 10, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter with software instruction to customers informing them about an upcoming software VB10B under(SY003/16/P) update change and issue that the new software version will address. For further questions, please call (610) 448-6478.
Quantity
74 systems
Distribution
US Distribution to the states of : AL, CA, CT, FL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, TX, WI
Lot, serial or model codes
Serial Numbers: 100389, 100198, 100258, 100259, 100538, 100133, 100213, 100214, 100031, 100484, 100362, 100233, 100140, 100141, 100142, 100143, 100144, 100416, 100105, 100106, 100109, 100110, 100111, 100112, 100113, 100114, 100115, 100116, 100117, 100119, 100120, 100457, 100245, 100447, 100364, 100314, 100298, 100299, 100386, 100139, 100027, 100300, 100197, 100025, 100084 100186, 100189, 100190, 100191, 100192, 100195, 100196, 100033, 100306, 100307, 100232, 100256, 100199, 100188, 100194, 100257, 100425, 100029, 100253, 100132, 100247, 100248, 100434, 100491, 100495, 100221, 100212
510(k) numbers
["K093612"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register