FDA Medical Device Recalls (openFDA)
Aesculap Implant Systems LLC: GP161R; ELAN 4 1-RING DIAMOND BURR COARSE D2.0; GP162R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2.0; GP163R; ELAN 4 1-RING DIAMOND BURR COARSE D2,3; GP164R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2,3; GP165R; ELAN 4 1-RING DIAMOND BURR COARSE D3,0; GP166R; ELAN 4 1-RING DIAMOND BURR+ COARSE D3,0; GP167R; ELAN 4 1-RING DIAMOND BURR++ COARSE D3,0; GP168R; ELAN 4 1-RING DIAMOND BURR COARSE D4,0; GP169R; ELAN 4 1-RING DIAMOND BURR COARSE D5,0; GP170R; ELAN 4 1-RING DIAMOND BURR COARSE D6,0; GP173R; ELAN 4 1-RING DIAMOND BURR X-COARSE D3,0; GP174R; ELAN 4 1-RING DIAMOND BURR X-COARSE D4,0; GP175R; ELAN 4 1-RING DIAMOND BURR X-COARSE D5,0; GP176R; ELAN 4 1-RING DIAMOND BURR X-COARSE D6,0; GP177R; ELAN 4 1-RING DIAMOND BURR X-COARSE D7,0; GP328R; ELAN 4 2-RING DIAMOND BURR X-COARSE D4.0; GP329R; ELAN 4 2-RING DIAMOND BURR X-COARSE D5.0; GP330R; ELAN 4 2-RING DIAMOND BURR X-COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
- Recall number
- Z-2087-2017
- Recalling firm
- Aesculap Implant Systems LLC
- Product
- GP161R; ELAN 4 1-RING DIAMOND BURR COARSE D2.0; GP162R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2.0; GP163R; ELAN 4 1-RING DIAMOND BURR COARSE D2,3; GP164R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2,3; GP165R; ELAN 4 1-RING DIAMOND BURR COARSE D3,0; GP166R; ELAN 4 1-RING DIAMOND BURR+ COARSE D3,0; GP167R; ELAN 4 1-RING DIAMOND BURR++ COARSE D3,0; GP168R; ELAN 4 1-RING DIAMOND BURR COARSE D4,0; GP169R; ELAN 4 1-RING DIAMOND BURR COARSE D5,0; GP170R; ELAN 4 1-RING DIAMOND BURR COARSE D6,0; GP173R; ELAN 4 1-RING DIAMOND BURR X-COARSE D3,0; GP174R; ELAN 4 1-RING DIAMOND BURR X-COARSE D4,0; GP175R; ELAN 4 1-RING DIAMOND BURR X-COARSE D5,0; GP176R; ELAN 4 1-RING DIAMOND BURR X-COARSE D6,0; GP177R; ELAN 4 1-RING DIAMOND BURR X-COARSE D7,0; GP328R; ELAN 4 2-RING DIAMOND BURR X-COARSE D4.0; GP329R; ELAN 4 2-RING DIAMOND BURR X-COARSE D5.0; GP330R; ELAN 4 2-RING DIAMOND BURR X-COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
- Status
- Terminated
- Initiated
- 2017-03-28
- Posted
- 2017-04-24
- Terminated
- 2018-02-06
- Reason
- Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
- Root cause
- Process control
- Action
- On April 7, 2017 8 Hospitals, 3 distributors and 11 sales reps were sent an Urgent Medical Device Recall Notification letter. Letters were sent UPS overnight. Customers were asked to immediately discontinue use and return the product to Aesculap, Inc.
- Quantity
- Total of all burrs - 608 units
- Distribution
- US Nationwide distribution - CA, IL, NY, PA, TN, TX
- Lot, serial or model codes
- Model numbers COARSE: GP161R ; GP162R ; GP163R ; GP164R ; GP165R ; GP166R ; GP167R ; GP168R ; GP169R ; GP170R ; Model numbers X-Course: GP173R ; GP174R ; GP175R ; GP176R ; GP177R ; GP328R ; GP329R ; GP330R ;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GEY
- Firm FEI
- 2916714
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28