Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics Inc.: Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge Module (10703032 or 10713761) FlexLab High Throughput Centrifuge Module (11171747) FlexLab and Aptio Centrifuge Module (10628157 or 11266581): Used to separate substances according to their densities.

Recall number
Z-2087-2016
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge Module (10703032 or 10713761) FlexLab High Throughput Centrifuge Module (11171747) FlexLab and Aptio Centrifuge Module (10628157 or 11266581): Used to separate substances according to their densities.
Status
Terminated
Initiated
2016-05-11
Posted
not on file
Terminated
2018-04-27
Reason
Premature bucket failures
Root cause
Under Investigation by firm
Action
Siemens mailed to customers an Urgent Medical Device Correction (UMDC#LAI16-02.A.US) Urgent Field Safety Notification letter informing customers of the problem with recommendations to follow the existing maintenance instructions in the product operator manuals as well as a new requirement to spin at cycle speeds not exceeding 3500 rpm.
Quantity
487
Distribution
Nationwide: AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, WI, WV.
Lot, serial or model codes
All serial numbers are affected.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXG
Firm FEI
3002755490
Firm city
Brookfield
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register