Skip to the content

FDA Medical Device Recalls (openFDA)

Northern Digital Inc.: NDI Disposable Reflective Marker Spheres for Brainlab IGS (Image Guided Surgical) system. Product Usage: Disposable reflective marker spheres are attached to reference arrays and instruments, thus enabling the infrared tracking systems to detect the position of the patient and instruments in the surgical field.

Recall number
Z-2084-2015
Recalling firm
Northern Digital Inc.
Product
NDI Disposable Reflective Marker Spheres for Brainlab IGS (Image Guided Surgical) system. Product Usage: Disposable reflective marker spheres are attached to reference arrays and instruments, thus enabling the infrared tracking systems to detect the position of the patient and instruments in the surgical field.
Status
Terminated
Initiated
2015-06-10
Posted
2015-07-16
Terminated
2016-10-03
Reason
Inadequate curing of adhesive, may cause the two halves of the spheres to separate.
Root cause
Process control
Action
NDI sent an Urgent Customer Notification letter dated June 10, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the Acknowledgement Form attached to the notification letter and return using the enclosed pre-paid FedEx label. For questions, contact NDI at 1 (877) 634-6340 ext. 544 (if calling from North America) or +1 (800) 634 634 00 ext. 544 (if calling from outside North America) .
Quantity
524, 700 spheres (41773 = 440,820 spheres; 41774 = 77,760 spheres and 41813 = 6,120 spheres)
Distribution
US Nationwide Distribution
Lot, serial or model codes
1) Catalog Number: 41773 (90 pack) All Lots beginning with the prefix C / All expiration dates; 2) Catalog Number: 41774 (270 pack) All Lots beginning with the prefix C / All expiration dates; and 3) Catalog Number: 41813 (60 pack) All Lots beginning with the prefix C / All expiration dates
510(k) numbers
["K100038"]
PMA numbers
not on file
Product code
HAW
Firm FEI
3002743211
Firm city
Waterloo Canada
Firm state
Ontario
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register