Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Tibial Punch Tower Assembly. Non-Sterile. CAT NO. 8000-1089. Howmedica Osteonics Corp., 59 Route 17, Allendale, NJ 07401. During primary Scorpio TKA procedures, the Scorpio Tibial Punch Tower (8000-1089) is used in conjunction with Scorpio Tibial Keel Punches (3760-0305/-0709/-1113 and 3761-0305/-0709/-1113) and Tibial Templates (3750-0003/-0005/-0007/-0009/-0013) to prepare the tibia for baseplate implantation. Per the surgical protocol for Duracon and Scorpio Total Knee Systems with Xcelerate Instrumentation (LSPK31 03/03), tibial templates are used to determine the appropriate coverage of the tibial plateau. Once a template size is selected, the device is secured to the plateau with fixation pins, and the tibial punch tower is assembled by placing the tower onto the two small locating pins on the tibial template. A keel punch is then selected, assembled into the channel of the tower, and impacted with a mallet. Depending on the intended implant size, several keel punches of increasing size may be utilized to incrementally enlarge the cavity. Each keel punch is advanced until it fully seats on the tibial template. Tibial preparation is considered complete when the final keel punch fully seats on the tibial template.

Recall number
Z-2084-2012
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Tibial Punch Tower Assembly. Non-Sterile. CAT NO. 8000-1089. Howmedica Osteonics Corp., 59 Route 17, Allendale, NJ 07401. During primary Scorpio TKA procedures, the Scorpio Tibial Punch Tower (8000-1089) is used in conjunction with Scorpio Tibial Keel Punches (3760-0305/-0709/-1113 and 3761-0305/-0709/-1113) and Tibial Templates (3750-0003/-0005/-0007/-0009/-0013) to prepare the tibia for baseplate implantation. Per the surgical protocol for Duracon and Scorpio Total Knee Systems with Xcelerate Instrumentation (LSPK31 03/03), tibial templates are used to determine the appropriate coverage of the tibial plateau. Once a template size is selected, the device is secured to the plateau with fixation pins, and the tibial punch tower is assembled by placing the tower onto the two small locating pins on the tibial template. A keel punch is then selected, assembled into the channel of the tower, and impacted with a mallet. Depending on the intended implant size, several keel punches of increasing size may be utilized to incrementally enlarge the cavity. Each keel punch is advanced until it fully seats on the tibial template. Tibial preparation is considered complete when the final keel punch fully seats on the tibial template.
Status
Terminated
Initiated
2011-12-22
Posted
2012-07-26
Terminated
2013-06-13
Reason
Stryker Orthopaedics received a report indicating that a Tibial Keel Punch could not be inserted into the slot of the Tibial Punch Tower. The slot that guides the stem of the keel punch into the channel of the tower was later determined to be undersized.
Root cause
Nonconforming Material/Component
Action
Stryker Orthopaedics sent an Urgent Product Recall/Recall Acknowledgement letter dated January 10, 2012, to all affected customers. The letter identified the product(s) the problem and the action needed to be taken by the customer. Customers were instructed to contact their Stryker Sales Representative to arrange for return of the affected product. Customers were asked to fax back the attached Product Recall Acknowledgement Form as soon as possible to 201-831-6069. For questions customers should call 201-972-2100. For questions regarding this recall call 201-831-5970.
Quantity
152 units
Distribution
Worldwide Distribution
Lot, serial or model codes
Catalog Number 8000-1089. Lot Number(s): RD6E087, RD6E187; RD6E083; RD6E082; RD6C178; RD5W268; RD5W267; RD5V260; RD5W320; RD5V240; RD5V239; RD5V238; RD5N407; RD5S104; RD5S103; RD5T177; RD5T179; RD5S259; RD5T178; RD5N406; RD5S300; RD5T298; RD5W381; RD6E086; RD6E089; RD6E090; RD6E091; RD6E158; RD6E159; RD6E160; RD6E186.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register