Medtronic Navigation, Inc.-Littleton: Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028

Recall number
Z-2082-2016
Recalling firm
Medtronic Navigation, Inc.-Littleton
Product
Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028
Status
Terminated
Initiated
2016-03-17
Posted
not on file
Terminated
2021-04-06
Reason
It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.
Root cause
Radiation Control for Health and Safety Act
Action
Medtronic's planned action to bring device into Compliance. 1. You will provide an Errata sheet to customers updating the missing and incorrect information. 2. Field Service Engineers (FSEs) will perform radiation output measurements via validated procedures to attain compliance with 21CFR part 1020.32(k)(6). 3. You will inspect and test X-ray generator and system motion batteries for damage, in accordance with the plan submitted to CDRH. Damaged batteries will be replaced. 4. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet. For further questions, please call (800) 595-9709.
Quantity
US - 581
Distribution
Nationwide Distribution
Lot, serial or model codes
Model No. BI-700-00027 and BT-700-00028
510(k) numbers
["K092564"]
PMA numbers
not on file
Product code
OXO
Firm FEI
3004785967
Firm city
Littleton
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register