Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Medical Care Holdings, Inc. dba Renal Solutions: 2008 MeDS Pump, Venofer(R) Pump Product Usage: The 2008 MeDs pump is an optional accessory for use with the Fresenius 2008 Series Hemodialysis Machines and is intended to administer Venofer to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer, in conjunction with hemodialysis, is prescribed by a physician.

Recall number
Z-2082-2013
Recalling firm
Fresenius Medical Care Holdings, Inc. dba Renal Solutions
Product
2008 MeDS Pump, Venofer(R) Pump Product Usage: The 2008 MeDs pump is an optional accessory for use with the Fresenius 2008 Series Hemodialysis Machines and is intended to administer Venofer to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer, in conjunction with hemodialysis, is prescribed by a physician.
Status
Terminated
Initiated
2013-04-01
Posted
2013-08-26
Terminated
2015-03-02
Reason
Product is being recalled because the combination of the current 50 mg vial, the 2008 MeDS pump, and the 2008 MeDS pump tubing Set for CombiSet Bloodlines can lead to a situation where Venofer(R) delivery does not meet dosage accuracy requirements. This may lead to a slight over or under delivery of Venofer(R) that does not pose an acceptable risk to the patient. Please note that this recall
Root cause
Device Design
Action
Renal Solutions sent an Urgent Notification letter dated April 1, 2013 via certified mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to cease further use of the device. The letter informed customers that a Fresenius Medical Care Technical Service/Customer Service representative will contact customers to arrange for removal of the device. For questions contact Fresenius Medical Care Customer Service at 800-227-2572.
Quantity
132
Distribution
USA Nationwide Distribution in the states of CA, MA, and NY.
Lot, serial or model codes
Serial numbers - TM-001065, TM-001283, TM-001284, TM-001325, TM-001411, TM-001426, TM-001465, TM-001476, TM-001529, TM-001967, TM-001968, TM-001975, TM-001985, TM-002015, TM-002016, TM-002017, TM-002018, TM-002021, TM-002022, TM-002095, TM-000015, TM-000028, TM-000031, TM-000041, TM-0048, TM-000050, TM-000051, TM-000053, TM-000055, TM-000059, TM-000064, TM-000065, TM-000068, TM-0069, TM-000213, TM-000215, TM-000218, TM-000219, TM-000070, TM-000072, TM-000074, TM-000075, TM-000078, TM-000081, TM-000082, TM-000083, TM-000085, TM-000086, TM-000087, TM-000088, TM-000089, TM-000092, TM-000040, TM-000077, TM-000098, TM-000101, TM-000102, TM-000103, TM-000104, TM-000105, TM-000106, TM-000107, TM-000108, TM-000109, TM-000112, TM-000113, TM-000114, TM-000115, TM-000116, TM-000122, TM-000123, TM-000124, TM-000125, TM-000126, TM-000128, TM-000145, TM-000002, TM-000003, TM-000004, TM-000005, TM-000006, TM-000007, TM-000008, TM-000009, TM-000010, TM-000012, TM-000013, TM-000014, TM-000017, TM-000018, TM-000019, TM-000020, TM-000023, TM-000024, TM-000025, TM-000033, TM-000036, TM-A595, TM-A610, TM-A627, TM-000220, TM-000227, TM-007364, TM-007371, TM-006848, TM-006860, TM-006875, TM-006879, TM-006880, TM-006881, TM-006882, TM-006883, TM-006884, TM-006885, TM-006886, TM-006887, TM-006890, TM-006891, TM-006892, TM-006893, TM-006897, TM-006898, TM-006911, TM-006912, TM-006913, TM-006914, TM-006915, TM-006916, TM-006917, TM-006918, TM-006919, and TM-006920.
510(k) numbers
["K121497"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3005778453
Firm city
Warrendale
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register