Hill-Rom, Inc.: Affinity Four Birthing Bed P3700 Product Usage: The Affinity, Four Birthing Bed is intended to be used as birthing beds for women of child bearing age in an LDR (Labor, Delivery, Recovery ) or LDRP (Labor, Delivery, Recovery, Postpartum) setting within the acute care Labor and Delivery market. They are not intended for use as general hospital beds.

Recall number
Z-2082-2012
Recalling firm
Hill-Rom, Inc.
Product
Affinity Four Birthing Bed P3700 Product Usage: The Affinity, Four Birthing Bed is intended to be used as birthing beds for women of child bearing age in an LDR (Labor, Delivery, Recovery ) or LDRP (Labor, Delivery, Recovery, Postpartum) setting within the acute care Labor and Delivery market. They are not intended for use as general hospital beds.
Status
Terminated
Initiated
2012-04-02
Posted
2012-07-26
Terminated
2015-07-21
Reason
During review of our last braking system via that field corrective action, the system still did not meet our expectations, which is to eliminate the hazards of patient falling. Bed movement immediately prior to, or during the birthing process may still create distractions and delays, resulting in patient or caregiver falls, or contribute to a deviation in a procedure, (e.g. IV insertion, epidural
Root cause
Device Design
Action
Hill-Rom sent an URGENT FIELD SAFETY NOTICE letter dated April 2, 2012 to all affected customers. The letter identified the affected products, problem and actions to be taken. Customers were informed that Hill-Rom will visit their facility to modify the affected products, however, until then follow the actions to be taken outlined. For questions concerning this request or the procedure as outlined contact Hill-Rom Technical Support at 800-445-3720 (U.S.).
Quantity
12,726 devices
Distribution
Worldwide Distribution - US Nationwide and countries including: Argentina , Aruba, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Guam, Guyana, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Malaysia, Mauritius, Mexico, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Turkey, Turks and Caicos Islands, United Arab Emirates, United Kingdom, United States, US Virgin Islands, Venezuela and Vietnam
Lot, serial or model codes
Affinity Four Birthing Beds (P3700) manufactured with Serial Numbers H331AA4785 thru N081AA7966
510(k) numbers
not on file
PMA numbers
not on file
Product code
KNC
Firm FEI
1824206
Firm city
Batesville
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register