Skip to the content

FDA Medical Device Recalls (openFDA)

Xcision Medical Systems, LLC: GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.

Recall number
Z-2081-2019
Recalling firm
Xcision Medical Systems, LLC
Product
GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.
Status
Terminated
Initiated
2019-07-03
Posted
not on file
Terminated
2021-07-01
Reason
One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.
Root cause
Nonconforming Material/Component
Action
The firm sent a Field Correction Notice letters dated 7/3/19 to customers. The letter identified the affected product, problem and actions to be taken. If you need any further information or support concerning this issue, please contact Xcsion Customer Support at 855-924-7466.
Quantity
2
Distribution
The products were distributed to the following US states: MD and TX.
Lot, serial or model codes
Serial Numbers GP0002, GP0004; UDI: (01) 00850002615005
510(k) numbers
["K180571"]
PMA numbers
not on file
Product code
IWB
Firm FEI
3014273653
Firm city
Columbia
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register