Ortho-Clinical Diagnostics: enGen (TM) Laboratory Automation Systems using all TCAutomation(TM) Software Versions with the InOut Communication Interface, IVD

Recall number
Z-2077-2017
Recalling firm
Ortho-Clinical Diagnostics
Product
enGen (TM) Laboratory Automation Systems using all TCAutomation(TM) Software Versions with the InOut Communication Interface, IVD
Status
Terminated
Initiated
2017-03-08
Posted
not on file
Terminated
2021-08-24
Reason
Software anomaly; Thermo-Fisher Scientific initially discovered and Ortho-Clinical Diagnostics, subsequently, confirmed a software anomaly that may potentially result in miss-associated sample IDs involving the communication between enGen TCAutomation (TCA) Bypass modules (manufactured by Thermo-Fisher Scientific) and the VITROS Analyzers. To date, no occurrences of this issue have been observed on Orthos enGen(TM) Systems. Sample IDs may become miss-associated with the actual sample being processed due to discrepant information passed to the analyzer through InOut communication device, triggered by a specific set of events occurring between the Bypass module and the analyzer. --- All of the following conditions must be present for the issue to potentially occur: (1.) The Bypass module is in the process of positioning Sample A at the aspiration position, and (2.) At least two additional sample tubes (Samples B and C) are waiting to be aspirated in the Bypass module queue, and (3.) A Radio Frequency Identification (RFID) Tag Read failure occurs on the Sample A carrier, and (4.) The VITROS System is not available to aspirate on the track (for example, the analyzer is initializing, or the user is making entries on the diagnostic screen, or samples are being aspirated from the front of the analyzer, etc.). --- If the software anomaly occurs and the issue is present, the Sample ID in the INOUT communication device software queue becomes asynchronous with the actual Sample ID physical queue, and the VITROS assay result will be reported for the wrong Sample ID. All subsequent samples will have miss-associated results until the Bypass module is reset or the queue is empty. --- This issue could potentially miss-associate a sample/patient ID with a test result of a different sample. Depending on the test involved, patients health condition, etc., the potential health risk posed to the affected patient could range from no risk to serious risk, and potentially could lead to misdiagnosis or inappropriate medical intervention. To date, no patient harm was reported related to this product issue.
Root cause
Software design
Action
On 3/8/2017, URGENT PRODUCT CORRECTION NOTIFICATION letter (Ref. CL2017-059, dated 3/8/2017) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all enGen customers to inform them of the issue and provide them instructions on how to monitor and detect if the issue has potentially occurred. For questions or additional information, please contact Ortho Care Technical Solutions Center at 1-800-421-3311. Foreign affiliates were informed by e-mail on 3/8/2017 of the issue and instructed to notify their consignees of the issue and appropriate actions to take.
Quantity
US: 33 units; OUS: 43 units
Distribution
MA, NJ, NY, NY, PA,DE, MD, WV, NC, GA, FL, TN, MS, OH, MI, IL, MO, NE, AR, TX, CO, ID, TX, CA and Foreign distribution to the following countries: Belgium, Bermuda, Brazil, Canada, Chile, Denmark, England, France, Italy, Mexico, Norway, Portugal, Romania, Singapore, Spain, Sweden, Taiwan, United Kingdom.
Lot, serial or model codes
US Serial Numbers/J Numbers: B3ZHP2J/JB3ZHP2J; 736BM4J/J736BM4J; 953289/J953289; 953252/J953252; HJ5K84J/JHJ5K84J; FCM073J/JFCM073J; HPKG25J/JHPKG25J; GWXQ52J/JGWXQ52J; GBX395J/GBX395J; JFBX395J/JFBX395J; FXCBG3J/JFXCBG3J; 3RHC23J/J3RHC23J; JG7BFB5J/JG7BFB5J; 953270/J953270; 953221/J953221; 54FC01J/J54FC01J; 953209/J953209; 5J4B45J/J5J4B45J; 5LQ6Q2J/ J5LQ6Q2J; 2Z3YV1J/J2Z3YV1J; FRKMB2J/JFRKMB2J; 5GV6W2J/5GV6W2J; 7KSC15J/J7KSC15J; 953276/J953276; J2QKG25J/J2QKG25J; DK2BG3J/JDK2BG3J; 953266/J953266; J86PWC5J/ J86PWC5J; JXBQW4J/JJXBQW4J; CXL941J/JCXL941J; J953224/J953224; 36PWC5J/J36PWC5J; 4HX7S4J/J4HX7S4J --- Foreign Serial Numbers/J #s: CRKM82/J31265; 2S5OP0J/J31067; C4Z655J/ J31421; 6VR7K4J/J31401;  90JYS22/J31350; HWJZY4J/JHWJZY4J; J3J4B45J/J3J4B45J; JCT20Q3J/JCT20Q3J; J953234/J953234; J41BCY4J/J41BCY4J; J953244/J953244; J1HX7S4J/ J1HX7S4J; J16XQH4J/J16XQH4J; GWJZY4J/ GWJZY4J; DWXO52J/DWXO52J; 17QP94J/J20022; BGV6W2J/ J20003; H5PWC5J/J271371; 2LSSF5J/J271377; 1K5K84J/ J271264; 8W7WG3J/J85206; 96XQHAJ/ J7023645; 1LSSF5J/J7025050; 86T205J/ J7024893; C3JTP4J/J7023847; D3JTP4J/J7023874; C1JNZ5J/J7023730;  7DRJ44J/J7023728; 7YCF43J/J7023435; 18BFB5J/J7025017; J953222/J953222; J2V6LW4J/J2V6LW4J; J56T205J/J56T205J; BGV6W2J/J57663; 4drj44j/J800103; 6KSC15J/J1801116; FWGFC4J/J75663;  7GV6W2J/J75615; BT9PB5J/ J75666; JT6LW4J/ J30372; 38769917923/JTAINAN1; F2250767/J3055503; F1150698/ J3055496.
510(k) numbers
["K063144"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register