FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: Access BR Monitor, Part No. 387620
- Recall number
- Z-2074-2017
- Recalling firm
- Beckman Coulter Inc.
- Product
- Access BR Monitor, Part No. 387620
- Status
- Terminated
- Initiated
- 2017-04-13
- Posted
- not on file
- Terminated
- 2018-06-04
- Reason
- Following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de scurit du Medicament et des produits de sant (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU).
- Root cause
- Other
- Action
- An urgent Medical Device Recall letter dated 4/13/17 was sent to customers following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de scurit du Medicament et des produits de sant (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU). Customers are informed of the impact and the actions to be taken. Customers with any questions regarding the notice are instructed to contact Customer Technical Support Center: From our website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter representative.
- Quantity
- 45,278 units total (2,419 units in US)
- Distribution
- Affected geographies: Algeria, Australia, Austria, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Venezuela, Vietnam, Zimbabwe
- Lot, serial or model codes
- 623913, 625701, 628771, 630349, 633070 and All Future Lots
- 510(k) numbers
- ["K033036","K072612"]
- PMA numbers
- not on file
- Product code
- MOI
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28