Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: Access BR Monitor, Part No. 387620

Recall number
Z-2074-2017
Recalling firm
Beckman Coulter Inc.
Product
Access BR Monitor, Part No. 387620
Status
Terminated
Initiated
2017-04-13
Posted
not on file
Terminated
2018-06-04
Reason
Following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de scurit du Medicament et des produits de sant (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU).
Root cause
Other
Action
An urgent Medical Device Recall letter dated 4/13/17 was sent to customers following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de scurit du Medicament et des produits de sant (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU). Customers are informed of the impact and the actions to be taken. Customers with any questions regarding the notice are instructed to contact Customer Technical Support Center: From our website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter representative.
Quantity
45,278 units total (2,419 units in US)
Distribution
Affected geographies: Algeria, Australia, Austria, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Venezuela, Vietnam, Zimbabwe
Lot, serial or model codes
623913, 625701, 628771, 630349, 633070 and All Future Lots
510(k) numbers
["K033036","K072612"]
PMA numbers
not on file
Product code
MOI
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register