TomoTherapy Incorporated: TomoTherapy Hi-Art System, ver. 4.0.0, 4.0.1, 4.0.2, ThomoTherapy Incorporated 1240 Deming Way, Madison, WI 53717. The TomoTherapy HI-ART System is intended to be used as an integrated system for the planning and delivery of intensity modulated radiation therapy (IMRT). The HI-ART System provides precise delivery of radiation to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.

Recall number
Z-2072-2010
Recalling firm
TomoTherapy Incorporated
Product
TomoTherapy Hi-Art System, ver. 4.0.0, 4.0.1, 4.0.2, ThomoTherapy Incorporated 1240 Deming Way, Madison, WI 53717. The TomoTherapy HI-ART System is intended to be used as an integrated system for the planning and delivery of intensity modulated radiation therapy (IMRT). The HI-ART System provides precise delivery of radiation to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.
Status
Terminated
Initiated
2010-06-23
Posted
2010-07-22
Terminated
2011-12-17
Reason
An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer. With a particular workflow, the Planning Station's Draft and/or final Plan Report can provide incorrect dose statistics.
Root cause
Software design
Action
Consignees were sent on 6/23/10 a TomoTherapy "Urgent Field Safety Notice Medical device Correction" letter dated June 22, 1010. The letter addressed the Issue, Product Affected, Recommended Action and Resolution. Prior to generating a draft or final Plan Report for a given plan, the plan must be closed and reopened at least once. This issue will be remedied by Hi-Art software version 4.0.3, scheduled for release in late 2010. Customers with questions were provided with a contact in the letter at the TomoTherapy Customer Interaction Center. The firm can be contacted at 608-824-2800.
Quantity
29
Distribution
CA, CO, IL, LA, MI, MO, NY, ND, OH, OK, PA,VA, WI, BELGIUM CANADA, KOREA, MALAYSIA, SWITZERLAND, ITALY, GREAT BRITAIN, FRANCE.
Lot, serial or model codes
Serial #'s:  110019, 110083, 110169, 110284, 110305, 110312, 110034, 110091, 110174, 110285, 110306, 110314, 110037, 110092, 110278, 110288, 110308, 110316, 110056, 110150, 110281, 110293, 110309, 110318, 110065, 110159, 110283, 110300, 110311
510(k) numbers
["K082005"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
3003873069
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register