TEI Biosciences, Inc.: SurgiMend PRS; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 10x15 cm Semi-Oval, 7x17 cm Fenestrated, 10x20 cm Fenestrated, 8x16 cm Semi-Oval, 15x15 cm Semi-Oval, 8x20 cm Fenestrated, 10x15 cm Slant Fenestrated, 7x17 cm Semi-Oval, 8x16 cm Thin Semi-Oval, 10x20 cm Thin Semi-Oval, 6x16 cm Semi-Oval, Thin Oval 8x12 cm, Thin Oval 10x15 cm, Meshed 20cm x 10cm

Recall number
Z-2070-2023
Recalling firm
TEI Biosciences, Inc.
Product
SurgiMend PRS; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 10x15 cm Semi-Oval, 7x17 cm Fenestrated, 10x20 cm Fenestrated, 8x16 cm Semi-Oval, 15x15 cm Semi-Oval, 8x20 cm Fenestrated, 10x15 cm Slant Fenestrated, 7x17 cm Semi-Oval, 8x16 cm Thin Semi-Oval, 10x20 cm Thin Semi-Oval, 6x16 cm Semi-Oval, Thin Oval 8x12 cm, Thin Oval 10x15 cm, Meshed 20cm x 10cm
Status
Open, Classified
Initiated
2023-05-23
Posted
2024-09-30
Terminated
not on file
Reason
Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing.
Root cause
Under Investigation by firm
Action
The firm sent recall notifications to customers via email on 5/23/2023 and by mail on 5/24/2023. The recall notification informs customers that Integra is conducting a removal of all Surgimend PRS, Surgimend PRS Meshed, Surgimend, Surgimend MP, Revize, Revize-X, PriMatrix Dermal Repair Scaffold, and PriMatrix AG Antimicrobial Dermal Repair Scaffold devices due to identified issues with in-process and finished goods endotoxin testing that may result in out of specification endotoxin results. Customers are to check their stock for any lot of products listed in the provided Appendix and immediately remove them from service. All customers are asked to complete the provided acknowledgement form, regardless if affected product is on hand, and return it by email to Integralife6562@sedgwick.com or by fax to 888-229-0320. Questions about the acknowledgement form can be directed to 888-656-6375. Customers are to keep a copy of the recall notice for their records. A prepaid shipping label has been provided to all consignees so that return of products may be facilitated. Returns are being accepted for credit, however if a refund is preferred the firm asks that thsi is indicated in the "special instructions" section of the acknowledgement form. If any affected devices have been implanted, customers are to monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol. Customers with any questions can reach out to their Integra Sales Representative or Customer Service at 1-800-654-2873.
Quantity
22,163 units
Distribution
Domestic: Nationwide Distribution.
Lot, serial or model codes
Item No. (UDI-DI): 606-004-100 (10381780112822), 606-004-101 (10381780112839), 606-004-102 (10381780112846), 606-004-103 (10381780112853), 606-004-104 (10381780112860), 606-004-105 (10381780112877), 606-004-106 (10381780112884), 606-004-107 (10381780112891), 606-004-108 (10381780112907), 606-004-109 (10381780112914), 606-004-110 (10381780112921), 606-005-001 (10381780112938), 606-005-002 (10381780112945), 606-007-001 (10381780357322).
510(k) numbers
["K171357","K071807"]
PMA numbers
not on file
Product code
FTM
Firm FEI
3004170064
Firm city
Boston
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register