Skip to the content

FDA Medical Device Recalls (openFDA)

Exactech, Inc.: Exactech Ambassador Cervical Plate System The Exactech Ambassador Cervical Place System is intended for anterior cervical fixation (C2-T1) for the following: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Recall number
Z-2070-2015
Recalling firm
Exactech, Inc.
Product
Exactech Ambassador Cervical Plate System The Exactech Ambassador Cervical Place System is intended for anterior cervical fixation (C2-T1) for the following: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.
Status
Terminated
Initiated
2015-05-21
Posted
2015-07-14
Terminated
2015-07-16
Reason
Shipped to a single consignee prior to completion of final inspection.
Root cause
Release of Material/Component prior to receiving test results
Action
Exactech contacted their consignee by telephone on May 21, 2015, of this recall for their two products and instructed him to recover the devices from the hospital. Devices were recovered from the hospital May 22, 2015, and returned to the firm on May 26. All devices have been recovered.
Quantity
4 devices.
Distribution
Nationwide Distribution to Florida only.
Lot, serial or model codes
Catalog Number 05-100-03-0042, Lot # 68122001.
510(k) numbers
["K142576"]
PMA numbers
not on file
Product code
KWQ
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register