FDA Medical Device Recalls (openFDA)
Pentax Medical Company: Colonoscope Intended to provide optical visualization of and therapeutic access to the Lower Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Large Bowel to the Cecum. The instrument is introduced via the rectum when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
- Recall number
- Z-2068-2016
- Recalling firm
- Pentax Medical Company
- Product
- Colonoscope Intended to provide optical visualization of and therapeutic access to the Lower Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Large Bowel to the Cecum. The instrument is introduced via the rectum when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
- Status
- Terminated
- Initiated
- 2016-04-06
- Posted
- 2016-06-28
- Terminated
- 2017-12-21
- Reason
- Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Root cause
- Device Design
- Action
- Pentax sent an Urgent Field Correction letter dated April 6, 2016 to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were reminded to carefully check that all accessories are intact, that no parts have fallen off and become lodged within the endoscope's instrument/suction channel and to ensure that any therapeutic devices passed through the instrument channel are accounted for after use. If the instrument/suction channel does become blocked or clogged due to the accumulation of debris, an accessory that cannot be removed, or other cause, do NOT attempt to correct the blockage or continue to use the endoscope. Customers were instructed to contact their local PENTAX Medical Service facility to have the endoscope repaired. Customers with questions were instructed to call 800-431-5880, ext 2064.
- Quantity
- 12,412 units
- Distribution
- Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
- Lot, serial or model codes
- Model #'s EC-3430L, EC-3430LK, EC-3470LK, EC-380LKP, EC-3830LK, EC-3930LZ, EC-3831L, EC-3832L, EC-3832LK, EC-3870LK, EC-3870LZK, EC-3872LK, FC-38LV, EC-3830TLK, EC-3832TL, EC-3870TLK, EC-3872TLK, EC-3440L, EC-3840K, EC-2990LI, EC-3490LI, EC-3490LK, EC-3490TLI, EC34-I10L, EC-3890LI, EC-3890LK, EC-3890TLK and EC38-I10L
- 510(k) numbers
- ["K131855","K951574","K951578","K951579","K961570"]
- PMA numbers
- not on file
- Product code
- FDF
- Firm FEI
- 1000080301
- Firm city
- Montvale
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28