Medline Industries Inc: Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.

Recall number
Z-2067-2019
Recalling firm
Medline Industries Inc
Product
Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.
Status
Terminated
Initiated
2019-05-10
Posted
2019-08-12
Terminated
2021-08-18
Reason
This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.
Root cause
Nonconforming Material/Component
Action
On May 10, 2019, the firm notified its distributor, Medline, of the recall. The distributor was asked to identify all inventory within its possession, and to cease further distribution. The distributor was asked to notify any customers to whom the recalled product had been distributed, and to instruct customers to cease use immediately. On May 13, 2019, Medline notified its customers of the recall. Medline's instructions to customers were as follow: 1. Immediately check your stock for the specific kit number and the affected kit lot number in the attached list. Quarantine any kits with the listed lot numbers. If you dont have any of the lots listed above skip to step 3. 2. Affix the enclosed stickers on the affected kits and return to stock. If additional stickers are needed please call 866-359-1704. Please be sure NOT to use the recalled Infusion Set components, pull another approved product from sterile supply. 3. Please fill out and return the enclosed Verification Form via email to recalls@medline.com or fax 866-767-1290. Customers with questions may call 1-866-359-1704.
Quantity
4890
Distribution
One distributor located in Illinois.
Lot, serial or model codes
lots 2018090590 2018112190 2018090690 2018101190 2018101590 2018121190 2019013190 2019030490 2019040190
510(k) numbers
not on file
PMA numbers
not on file
Product code
N/A
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register