Skip to the content

FDA Medical Device Recalls (openFDA)

Windstone Medical Packaging, Inc.: Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F

Recall number
Z-2066-2026
Recalling firm
Windstone Medical Packaging, Inc.
Product
Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F
Status
Open, Classified
Initiated
2026-04-02
Posted
2026-05-08
Terminated
not on file
Reason
Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.
Root cause
Device Design
Action
Firm began notifying consignees via email on April 2, 2026. Notice was titled "URGENT: MEDICAL DEVICE RECALL." Consignees were instructed to check inventory and quarantine affected products. Firm instructed consignees to over-label affected products with a sticker which states that at the time the kit is opened, the affected Medline syringe should be identified and set aside, and that the affected syringe should be rendered unusable and disposed of. Firm will provide labels for any kits that need to be identified.
Quantity
150 kits (US only)
Distribution
US Nationwide distribution in the state of Arizona.
Lot, serial or model codes
Pack AMS6908E UDI-DI B098AMS6908E0 Lot 214875 Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRO
Firm FEI
1000125955
Firm city
Billings
Firm state
MT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register