Skip to the content

FDA Medical Device Recalls (openFDA)

Waldemar Link GmbH & Co. KG (Mfg Site): LinkSymphoKnee Distal Femoral Augment - Tilastan - Size 7-8, Medial-Right/Lateral Left, 5mm- Intended for primary and revision total knee replacement in skeletally mature patients. Product Code: 880-307/11

Recall number
Z-2066-2023
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product
LinkSymphoKnee Distal Femoral Augment - Tilastan - Size 7-8, Medial-Right/Lateral Left, 5mm- Intended for primary and revision total knee replacement in skeletally mature patients. Product Code: 880-307/11
Status
Open, Classified
Initiated
2022-12-23
Posted
2023-07-03
Terminated
not on file
Reason
Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).
Root cause
Component change control
Action
LinkBio Corporation issued A Product Advisory Notice to consignees holding field inventory on 18-May-2023. The product is provided to the surgeon directly by the representative of the Distributor. Letter states reason for recall, health risk and action to take: Please ensure the surgical team is made aware of this issue before using one of the affected augments. While there is no safety impact due to this issue, there could be some confusion during assembly if the issue is not known in advance. No additional action is required by users beyond reinserting the screw in the correct orientation, and no product return is required at this time. Please contact your LinkBio Sales or Customer Service representative or the contact for any questions: Jennifer Lubrecht Senior Director, Quality Management LinkBio Corporation 69 King Street Dover, NJ 07801 Mobile: +1 862.477.0370 Email: j.lubrecht@LinkBio.com
Quantity
20 units
Distribution
US Nationwide distribution in the states of AL, IL, IN, LA, GA, NV, TX.
Lot, serial or model codes
UDI-DI: 04026575257379 Lot Number: 2123225
510(k) numbers
["K202924"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3003386935
Firm city
Norderstedt
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register