Erbe Medical, LLC: ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222
- Recall number
- Z-2064-2026
- Recalling firm
- Erbe Medical, LLC
- Product
- ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222
- Status
- Open, Classified
- Initiated
- 2026-03-31
- Posted
- 2026-05-01
- Terminated
- not on file
- Reason
- Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
- Root cause
- Component change control
- Action
- On 3/31/2026, correction notices were emailed to Medical Directors, Risk Managers, Medical Device Safety Officers, or Healthcare Professionals informing them to take the following actions: 1) Discontinue use of unsupported third-party Scope Port Connectors. 2) Provide this notice to all personnel within your organization who use, service, purchase, stock, or otherwise manage. 3) If you further distributed any affected devices, promptly forward this notice to those accounts and request that they take the same actions. 4) Continue following the instructions for use, including verifying connector compatibility, ensuring that all connectors are fully seated, confirming that all joints and interfaces are secure, and verifying correct placement and alignment of the tubing within the pump head. 5) Utilize the compatibility table in the notice until updated Notes on Use is received. 6) Complete and return the response form via email to fieldaction.medical.us@erbegroup.com. Firm contacts: Correction: fieldaction.medical.us@erbegroup.com, 480-564-1899. Routine Customer Service and Complaints: techservices@erbegroup.com, 1-800-778-3723
- Quantity
- 5,500
- Distribution
- US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
- Lot, serial or model codes
- UDI-DI: 04065655000252. Lots: W4465580, W4466022, W4466023, WO469442, WO469923, WO473978, WO473979, WO473980, WO473981, WO476347, WO477013, WO477014
- 510(k) numbers
- ["K190469"]
- PMA numbers
- not on file
- Product code
- OCX
- Firm FEI
- 3020712480
- Firm city
- Tempe
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28