Fenwal Inc: VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to be used in healthcare facilities environment by healthcare trained professionals according to hospital protocols on adults, pediatrics and neonatal human patient to administer via a single channel or mounted on a multiple channels rack accessory: Intermittent or continuous delivery of parenteral fluids (solutions, colloids, parenteral nutrition) and medications (including but limited to diluted drugs, chemotherapy) through clinically accepted IV routes of administration). Transfusion of blood and blood derivatives products. Note that for transfusion procedures, use of a special administration set is required. Volumat MC Agilia can be used with the Dose Error Reduction Software (DERS) Vigilant Drug Lib to define the default and upper hard limits for the Bolus Volume, or dose, by drug. The health care professional can then operate the pump for that drug within the upper hard limits defined in the Drug Library.
- Recall number
- Z-2064-2019
- Recalling firm
- Fenwal Inc
- Product
- VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to be used in healthcare facilities environment by healthcare trained professionals according to hospital protocols on adults, pediatrics and neonatal human patient to administer via a single channel or mounted on a multiple channels rack accessory: Intermittent or continuous delivery of parenteral fluids (solutions, colloids, parenteral nutrition) and medications (including but limited to diluted drugs, chemotherapy) through clinically accepted IV routes of administration). Transfusion of blood and blood derivatives products. Note that for transfusion procedures, use of a special administration set is required. Volumat MC Agilia can be used with the Dose Error Reduction Software (DERS) Vigilant Drug Lib to define the default and upper hard limits for the Bolus Volume, or dose, by drug. The health care professional can then operate the pump for that drug within the upper hard limits defined in the Drug Library.
- Status
- Terminated
- Initiated
- 2019-06-24
- Posted
- 2019-08-12
- Terminated
- 2023-03-06
- Reason
- The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib
- Root cause
- Other
- Action
- The firm initiated the recall by letter on 06/24/2019. The letter describes the issues and provided supplemental instructions to be implemented until the new software is installed. The firm issued a press release on 06/28/2019.
- Quantity
- 9461 units
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- All serial numbers distributed in the US
- 510(k) numbers
- ["K121613"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 3004548776
- Firm city
- Lake Zurich
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28