Skip to the content

FDA Medical Device Recalls (openFDA)

C.R. Bard, Inc.: Bardex Lubricath 3-Way 75cc Continuous Irrigation Foley Catheter (Creevy), 22Fr. Rx (01267522)

Recall number
Z-2064-2018
Recalling firm
C.R. Bard, Inc.
Product
Bardex Lubricath 3-Way 75cc Continuous Irrigation Foley Catheter (Creevy), 22Fr. Rx (01267522)
Status
Terminated
Initiated
2017-12-13
Posted
not on file
Terminated
2020-12-14
Reason
Foley Catheter with Incorrect Retainer Cap Print
Root cause
Process control
Action
On December 13, 2017 a letter was sent to all their customers informing them of the following: Bard Medical Division (BMD), C.R Bard, Inc., would like to inform you of a potential labeling issue involving the BARD product code and lot numbers listed below. Manufacturing of these affected products began May 2017.Due to a manufacturing issue, there is a potential for catheters with the two lot numbers listed above to have the incorrect information printed on the purple retainer cap. The incorrect print specifies that the balloon should be inflated with 10mL of water. However, the correct volume to use to inflate the catheter balloon is 75cc . If the user underinflates the balloon with only 10mL of water, the expected effect of helping stop the bleeding as an adjunct to conventional hemostasis post-TURP may not happen as quickly as desired, likely leading to a marginal severity of harm. Additionally, the catheter may become dislodged from the patient. Instructions to Customer: "Examine your inventory and identify any product subject to this communication. During use, adhereto the correct inflation volume as stated in the packaging. "Product return is not required. If a unit does not meet your needs or you have any questions,please contact our customer service at 1-800-526-4455. "If you have further distributed or transferred this product, please identify the respective organizationsand notify them at once of this product communication. Your notification to these organizations maybe enhanced by including a copy of this medical device communication letter. We appreciate your cooperation and assistance in dealing with this matter and sincerely apologize for any inconvenience that may result from this action. If you have any questions please do not hesitate to contact your Bard representative.
Quantity
not on file
Distribution
FL, NY, MN and TX Canada, Japan and Taiwan
Lot, serial or model codes
Catalog # 01267522 Lot #'s: NGBR0383 and NGBR2129
510(k) numbers
["K910195"]
PMA numbers
not on file
Product code
EZC
Firm FEI
1018233
Firm city
Covington
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register