FDA Medical Device Recalls (openFDA)
Philips Medical Systems Nederlands: S8-3t Micro TEE Transducer is an Ultrasonic Diagnostic. The Part # is 989605379471. The User information is labeled in part: "***iE33 Ultrasound System ***User Information Update***4535 614 04811 Rev A***New S8-3t TEE Transducer*** This release of the iE33 Ultrasound System supports the S8-3t transesophageal echocardiographic (TEE) transducer***Manufactured by Philips Ultrasound***22100 Bothell-Everett Highway***Bothell, WA 98021-8431***USA***Telephone: +1 425-487-7000 or 800-426-2670***Fax: +1 425-485-6080***www.philips.com/ultrasound***". The S8-3t transducer is indicated for adult cardiac, pediatric cardiac and transesophageal uses, and operates in the following modes: B, M, Color M, CWD, PWD, high PRF, color Doppler, harmonic, XRES. The S8-3t Micro TEE is indicated for use on patients weighing at least 2.5 kg undergoing echocardiography exams in the ICU, echocardiography lab, interventional suite, or in the operating room during cardiac surgery. The S8-3t Micro TEE transducer can be identified by the designation "S8-3t" printed on the connector housing of the transducer.
- Recall number
- Z-2062-2011
- Recalling firm
- Philips Medical Systems Nederlands
- Product
- S8-3t Micro TEE Transducer is an Ultrasonic Diagnostic. The Part # is 989605379471. The User information is labeled in part: "***iE33 Ultrasound System ***User Information Update***4535 614 04811 Rev A***New S8-3t TEE Transducer*** This release of the iE33 Ultrasound System supports the S8-3t transesophageal echocardiographic (TEE) transducer***Manufactured by Philips Ultrasound***22100 Bothell-Everett Highway***Bothell, WA 98021-8431***USA***Telephone: +1 425-487-7000 or 800-426-2670***Fax: +1 425-485-6080***www.philips.com/ultrasound***". The S8-3t transducer is indicated for adult cardiac, pediatric cardiac and transesophageal uses, and operates in the following modes: B, M, Color M, CWD, PWD, high PRF, color Doppler, harmonic, XRES. The S8-3t Micro TEE is indicated for use on patients weighing at least 2.5 kg undergoing echocardiography exams in the ICU, echocardiography lab, interventional suite, or in the operating room during cardiac surgery. The S8-3t Micro TEE transducer can be identified by the designation "S8-3t" printed on the connector housing of the transducer.
- Status
- Terminated
- Initiated
- 2011-03-27
- Posted
- 2011-04-25
- Terminated
- 2012-02-14
- Reason
- Users of the Philips model S8-3t Micro TEE Transducer have reported that significant image quality degradation may occur after a number of use/cleaning cycles. If users do not inspect the transducer for such damage prior to each use, imaging may not always be available, which could pose a risk for patients.
- Root cause
- Other
- Action
- Philips Healthcare began mailing Urgent Medical Device Correction letter on March 28, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers are being instructed on how to inspect the distal tip for defects that cause image quality degradation, and to plan for epicardial / transthoracic ultrasound imaging as an adjunct to the S8-3t. Customers are also reminded to carefully follow cleaning described under Handling the S8-3t Transducer, of Philips' iE33 System User Information. Customers can call Philips Customer Service at 800-722-9377 and reference to FCO 79500198 for any questions about this recall.
- Quantity
- 155 units
- Distribution
- Worldwide Distribution - USA AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KY, LA, MA, MD, MI, MN, MS, NC, ND, NH, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI, and the countries of Australia, Austria, Belgium, Chile, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Korea Rep, Netherlands, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Ukraine, and United Kingdom.
- Lot, serial or model codes
- System Code Serial Numbers: 02KHRT, 02L66Z, 02LJ0G, 02LJKW, 02R6K9, 02R8HK, 02R8JJ, 02RB69, 02RBRM, 02RBVK, 02RDQJ, 02RDTX, 02RPDG, 02RPT9, 02X3KY, 02X3RG, 02X3WD, 02X3WD, 02X434, 02X43J, 02X43Y, 02X4JH, 02XDNH, 02XF24, 02XF6J, 02XJXG, 02XK9H, 02XKBF, 02XKCH, 02XKCQ, 02XKCQ, 02XKCY, 02XL54, 02XRN8, 02XT0Z, 02XTHZ, 034JBN, 034K6R, 034KC0, 034L0J, 034L34, 034L4W, 034L5X, 034LDN, 034V9K, 034VL2, 034VMJ, 034VP2, 034VTC, 034VZ0, 034W05, 034W20, 0353HK, 039VGF, 039VPZ, 039VWY, 039VZL, 039W1G, 039W66, 039W86, 039W9X, 039WB4, 039WB8, 039WBY, 039WX8, 039WY1, 039WY2, 039X41, 039X49, 039X99, 039XC1, 039XHG, 039XHG, 03B48J, 03B49T, 03B4GG, 03B4H4, 03B4HC, 03B4JQ, 03B4QR, 03B4V6, 03B4ZX, 03B51R, 03B521, 03B524, 03B566, 03B57R, 03B597, 03B8QB, 03B8RR, 03B8TX, 03B8TZ, 03B8V3, 03B8X1, 03B8X6, 03B8ZY, 03B905, 03B906, 03B90L, 03B912, 03B92H, 03B92H, B008YN, B008YY, B00905, B00919, B009BW, B009FG, B009FY, B009JX, B009MY, B009VM, B009ZW, B009ZW, B00D88, B00DBN, B00DF1, B00DL4, B00DLC, B00DQG, B00DR7, B00DXP, B00DZX, B00F05, B00F11, B00F14, B00F20, B00F3C, B00F5M, B00F5V, B00F61, B00F7J, B00F89, B00HRV, B00J3X, B00J73, B00JCD, B00JG5, B00JK0, B00JKG, B00JMC, B00P7K, B00P7R, B00P9C, B00PB0, B00PDV, B00PFB, B00PGV, B00PKG, B00PM0, B00PMP, B00PPW, B00PQ5, B00PT5, and B00PW4.
- 510(k) numbers
- ["K030455"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 3002807880
- Firm city
- Best
- Firm state
- not on file
- Firm country
- Netherlands
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28