Shimadzu Medical Systems: Sonialvision Safire II, Model #: DAR-8000f Product Usage: This angiographic x-ray system device intended to be used for the radiography in the hospital with X-ray devices. This device is operated and used by the physicians and X-ray technologist. The object of this device is total patient population.

Recall number
Z-2060-2018
Recalling firm
Shimadzu Medical Systems
Product
Sonialvision Safire II, Model #: DAR-8000f Product Usage: This angiographic x-ray system device intended to be used for the radiography in the hospital with X-ray devices. This device is operated and used by the physicians and X-ray technologist. The object of this device is total patient population.
Status
Terminated
Initiated
2018-01-12
Posted
not on file
Terminated
2022-02-24
Reason
When selecting serial radiography with a pulse rate of 7.5fps (including selecting a preset or changing the pulse rate using a temporary edit function) it was observed the indicated "irradiation time" on the X-ray Generator Console, as well as the "integral dose" on the external console of fluoroscopy, were reset to 0. There has been one report of this event.
Root cause
Software design
Action
Shimadf0u sent an Urgent Voluntary Medical Device Recall Notice letter dated January 2018. The letter identified the affected product, problem and actions to be taken. For questions contact Shimadzu representative or contact Shimadzu National Technical Support at 800-228-1429.
Quantity
17
Distribution
US Nationwide in the states of IL, WA, MI, TX, AZ, CA, LA, FL, SC, and NJ.
Lot, serial or model codes
Serial #: 0161G68101, 0261G65303, 0261G65702, 0261G68204, 3M72BFB21001, 3M72BFB2A001, 3M72BFC2A001, 3M72BFC2C001, 3M72BFB32001, 3M72BFB32006, 3M72BFB33003,3M72BFC33001, 3M72BFC34001, 3M72BFB35005, 3M72BFC37002, 3M72BFC42001, 3M72BFB46002
510(k) numbers
["K052500"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1000125362
Firm city
Torrance
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register