FDA Medical Device Recalls (openFDA)
Shimadzu Medical Systems: Sonialvision Safire II, Model #: DAR-8000f Product Usage: This angiographic x-ray system device intended to be used for the radiography in the hospital with X-ray devices. This device is operated and used by the physicians and X-ray technologist. The object of this device is total patient population.
- Recall number
- Z-2060-2018
- Recalling firm
- Shimadzu Medical Systems
- Product
- Sonialvision Safire II, Model #: DAR-8000f Product Usage: This angiographic x-ray system device intended to be used for the radiography in the hospital with X-ray devices. This device is operated and used by the physicians and X-ray technologist. The object of this device is total patient population.
- Status
- Terminated
- Initiated
- 2018-01-12
- Posted
- not on file
- Terminated
- 2022-02-24
- Reason
- When selecting serial radiography with a pulse rate of 7.5fps (including selecting a preset or changing the pulse rate using a temporary edit function) it was observed the indicated "irradiation time" on the X-ray Generator Console, as well as the "integral dose" on the external console of fluoroscopy, were reset to 0. There has been one report of this event.
- Root cause
- Software design
- Action
- Shimadf0u sent an Urgent Voluntary Medical Device Recall Notice letter dated January 2018. The letter identified the affected product, problem and actions to be taken. For questions contact Shimadzu representative or contact Shimadzu National Technical Support at 800-228-1429.
- Quantity
- 17
- Distribution
- US Nationwide in the states of IL, WA, MI, TX, AZ, CA, LA, FL, SC, and NJ.
- Lot, serial or model codes
- Serial #: 0161G68101, 0261G65303, 0261G65702, 0261G68204, 3M72BFB21001, 3M72BFB2A001, 3M72BFC2A001, 3M72BFC2C001, 3M72BFB32001, 3M72BFB32006, 3M72BFB33003,3M72BFC33001, 3M72BFC34001, 3M72BFB35005, 3M72BFC37002, 3M72BFC42001, 3M72BFB46002
- 510(k) numbers
- ["K052500"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 1000125362
- Firm city
- Torrance
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28