Skip to the content

FDA Medical Device Recalls (openFDA)

Savaria Concord Lifts, Inc.: Savaria SL-1000 Stairlift

Recall number
Z-2060-2017
Recalling firm
Savaria Concord Lifts, Inc.
Product
Savaria SL-1000 Stairlift
Status
Terminated
Initiated
2017-04-12
Posted
not on file
Terminated
2018-03-05
Reason
The electrical controller, combined with motor, causes feedback, damaging a component on the control board, starting a fire.
Root cause
Component design/selection
Action
The firm sent a letter to dealers dated July 26, 2016. The firm will contact consignees (distributors) via e-mail and phone.
Quantity
1425 in the United States 734 in Canada
Distribution
Distribution in the United States and Canada
Lot, serial or model codes
s/n or lot number - Job Number lots 671494 - 672668
510(k) numbers
not on file
PMA numbers
not on file
Product code
PCD
Firm FEI
3005753836
Firm city
Brampton
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28