Skip to the content

FDA Medical Device Recalls (openFDA)

Mindray DS USA, Inc. dba Mindray North America: V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.

Recall number
Z-2059-2014
Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Product
V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.
Status
Terminated
Initiated
2014-05-19
Posted
2014-07-16
Terminated
2017-04-20
Reason
ECG interpretation feature issues. False measurements reported when the algorithm cannot determine P wave or QRS axis. The Median Complex Report displays the representative beat 25% larger than actual signal input. When a patient date of birth is imported from an external source, the ECG interpretation algorithm uses the default age of 50, and interpretive statements may be incorrect.
Root cause
Software design
Action
Mindray issued "Urgent: V-Series Monitoring System Corrective Action" notices dated May 19, 2014, to customers that received the affected product via certified mail with return receipt verification. The notification informs the customer of the issues with the affected product and how to identify the affected product. The notification recommends that customers discontinue use of the 12-lead ECG interpretation feature until a software correction is available (anticipate August 2014). A point of contact is provided to customers in case they have questions. For further questions please call 1-800-288-2121.
Quantity
113 units (57 United States; 56 International)
Distribution
Worldwide Distribution - US including the states of Louisiana, New Jersey, Montana, South Carolina and Wyoming.
Lot, serial or model codes
Part Number (P/N): 0998-00-1800-101 and 0998-00-1800-201 with software version 2.4.1.18 in use with 12-lead ECG Module, P/N:0998-00-1804-01 and 0998-00-1804-02.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MHX
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register