FDA Medical Device Recalls (openFDA)
Mindray DS USA, Inc. dba Mindray North America: V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.
- Recall number
- Z-2059-2014
- Recalling firm
- Mindray DS USA, Inc. dba Mindray North America
- Product
- V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.
- Status
- Terminated
- Initiated
- 2014-05-19
- Posted
- 2014-07-16
- Terminated
- 2017-04-20
- Reason
- ECG interpretation feature issues. False measurements reported when the algorithm cannot determine P wave or QRS axis. The Median Complex Report displays the representative beat 25% larger than actual signal input. When a patient date of birth is imported from an external source, the ECG interpretation algorithm uses the default age of 50, and interpretive statements may be incorrect.
- Root cause
- Software design
- Action
- Mindray issued "Urgent: V-Series Monitoring System Corrective Action" notices dated May 19, 2014, to customers that received the affected product via certified mail with return receipt verification. The notification informs the customer of the issues with the affected product and how to identify the affected product. The notification recommends that customers discontinue use of the 12-lead ECG interpretation feature until a software correction is available (anticipate August 2014). A point of contact is provided to customers in case they have questions. For further questions please call 1-800-288-2121.
- Quantity
- 113 units (57 United States; 56 International)
- Distribution
- Worldwide Distribution - US including the states of Louisiana, New Jersey, Montana, South Carolina and Wyoming.
- Lot, serial or model codes
- Part Number (P/N): 0998-00-1800-101 and 0998-00-1800-201 with software version 2.4.1.18 in use with 12-lead ECG Module, P/N:0998-00-1804-01 and 0998-00-1804-02.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MHX
- Firm FEI
- 2221819
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as