Philips Healthcare Inc.: Philips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.

Recall number
Z-2059-2013
Recalling firm
Philips Healthcare Inc.
Product
Philips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.
Status
Terminated
Initiated
2013-08-08
Posted
2013-08-26
Terminated
2020-09-03
Reason
The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
Root cause
Software design
Action
Philips Healthcare sent an Urgent Field Safety Notice letter dated July 24, 2013, to all affected customer. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed : When reading an image, a physician should not use the mirror icon for image side interpretation. 2. Clinical staff should follow the Instructions for Use and use lead letters to indicate body orientation of patient or equivalent electronic markers to provide Left/Right indications, during all X-ray image acquisitions. Philips will upgrade the system, which involves software and hardware upgrades of the system. The software provides a replacement icon that reduces the risk of the potential misinterpretation of the original mirror icon. Philips will also provide updated Instructions for Use for the work spot and the system (including an addendum). If you need any further information or support concerning this, please contact your local Philips representative Technical Support Line at 1-800-722-9377. We apologize for any inconvenience this may cause and trust that this information is adequately addressing any concerns you may have.
Quantity
89
Distribution
Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, GA, MO, MS, NC, NH, NY, PA and TX., and the countries of Canada, Australia, China, Czech Republic, France, Germany, Greece, Italy, Netherlands, Portugal, Russia, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates and United Kingdom.
Lot, serial or model codes
437765/SN09000009, 452854/SN10000003, 463879/SN10000016, 500643/SN12000006,504710/SN12000011, 511454/SN12000018, 519869/SN13000005, 524432/SN13000007, 500643/SN12000006, 436570/SN09000008, 511492/SN12000019, SN09020039, SN09010077,  438946/SN09000011, SN10002064, SN11020126, SN12020282, 516048/SN12000021, SN13020163, 524411/SN13000011, 409205/SN08000002, 438474/SN09000040, 448425/SN10000001, 449743/SN10000010, 461559/SN10000014, 493489/SN11000008, 496231/SN12000003, 434381/SN09000005, 456026/SN10000011, 465845/SN11000005, 471608/SN11000001, 481912/SN11000004, 508150/SN12000015, 524367/SN13000009, 524510/SN13000010, SN07020246, 485916/SN11000007, 428148/SN09000006, 423585/SN09000003, 500351/SN12000009, 506215/SN12000012, 520201/SN13000003, 409442/SN08000003, 419279/SN09000001, 518054/SN13000001, 460532/SN10000013, SN09020304, 443304/SN09000012, 443310/SN09000013, 416734/SN08000005, 420531/SN09000002, 423195/SN09000004, 460544/SN10000017, 454997/SN10000009, 496323/SN12000002, 496324/SN12000001, 399965/SN08000001, 438232/SN09000010, SN09020302, 443840/SN09000014, 452406/SN10000004, 454310/SN10000005, 478444/SN11000003, 502350/SN12000007, 502355/SN12000016, 520759/SN13000006, 434429/SN09000007, 451666/SN10000002, 453781/SN10000008, 453960/SN10000007, 459266/SN10000012, 460755/SN10000015,  466174/SN10000019, 468076/SN10000020, 471702/SN11000002, 499486/SN12000004, 501950/SN12000008, 503132/SN12000010, 507693/SN12000013, 507694/SN12000014, 519994/SN13000004, 520759/SN13000006,SN09001033, SN09020312, SN09000826, SN11000761, 493873/SN11000009, SN12020117, SN12020322, SN12020440
510(k) numbers
["K982795"]
PMA numbers
not on file
Product code
KPR
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register