FDA Medical Device Recalls (openFDA)
GE Healthcare, LLC: SIGNA Creator, SIGNA Explorer, Nuclear Magnetic Resonance Imaging System
- Recall number
- Z-2058-2021
- Recalling firm
- GE Healthcare, LLC
- Product
- SIGNA Creator, SIGNA Explorer, Nuclear Magnetic Resonance Imaging System
- Status
- Open, Classified
- Initiated
- 2021-06-04
- Posted
- 2021-07-12
- Terminated
- not on file
- Reason
- GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.
- Root cause
- Software design
- Action
- Deployment of the original customer letter to consignees in the U.S. began on June 4, 2021. Dissemination of the updated customer letter to consignees in the U.S. began on June 11, 2021. The letters discuss an issue on the affected products listed below where the Patient Orientation button may inadvertently be clicked when intending to click on the Save RX button. Users are directed to review and confirm that the actual patient orientation matches the prescribed patient orientation before initiating the scan. A correction is planned.
- Quantity
- 1737 devices
- Distribution
- Worldwide distributions.
- Lot, serial or model codes
- N/A
- 510(k) numbers
- ["K143251"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28