Zeiss, Carl Inc: Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple Laser Scanning Microscopes. Laser scanning microscopes are used in cell biology research.

Recall number
Z-2058-2018
Recalling firm
Zeiss, Carl Inc
Product
Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple Laser Scanning Microscopes. Laser scanning microscopes are used in cell biology research.
Status
Terminated
Initiated
2018-05-08
Posted
2018-06-09
Terminated
2020-04-21
Reason
In certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit.
Root cause
Radiation Control for Health and Safety Act
Action
1.Your firm will send an initial email to all known users advising them of the issue, the risk involved and associated mitigations until the units can be evaluated by Carl Zeiss service representatives, 2.Your firm will send a second written communication to each first line purchaser and to each known user site with the notice of action and warning to discontinue use of the component until it has been verified as defective or properly functioning, and 3.Your firm will dispatch factory trained service representatives to perform verification tests and to replace defective units as necessary, free of charge. CDRH approves the CAP subject to the following conditions: For further questions, please call (914) 747-1800.. 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Corrective actions will be provided at no cost to the purchasers and completed by March 1, 2019.
Quantity
1070 units to US
Distribution
USA (nationwide)
Lot, serial or model codes
Model No. Cell Observer SD, Laser TIRF, and DirectFRAP Laser Scanning Microscopes
510(k) numbers
not on file
PMA numbers
not on file
Product code
REM
Firm FEI
2431026
Firm city
Thornwood
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register