Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew Inc: R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 62 MM O.D., REF 71338962, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06

Recall number
Z-2058-2011
Recalling firm
Smith & Nephew Inc
Product
R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 62 MM O.D., REF 71338962, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
Status
Terminated
Initiated
2011-03-11
Posted
2011-04-22
Terminated
2012-02-08
Reason
During the manufacturing process for several batches of R3 Ceramic Liners, the titanium rings were pressed onto the ceramic component with a higher force than allowed by manufacturing specifications. This has the potential to result in lower than expected strength for the liners.
Root cause
Employee error
Action
All affected Smith & Nephew Distributors were notified via email and the hospitals were notified by letter delivered by Fed-Ex on 3/11/2011. They were instructed to immediately cease distribution, quarantine affected product for return, if product further distributed, identify customers and immediately notify them of recall. Please contact Mr. David Archer, Group Director, at 701.396.2121, ext. 5373, if you have questions.
Quantity
67 units
Distribution
AK, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SC, TN, TX, WA, and WI
Lot, serial or model codes
09JT35383
510(k) numbers
not on file
PMA numbers
["P030022"]
Product code
MRA
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register