Med-Hot Thermal Imaging, Inc.: MEDHOT MTI 2000 THERMAL IMAGING SYSTEM - TotalVision software with Elevated Body Temperature Screening module, for use as a medical diagnostic purposes as part of a telethermographic system utilizing FLIR cameras, Models A615sc and A315sc. Cameras are private labeled as MAX Basic, MAX 76 and MAX 307.

Recall number
Z-2057-2017
Recalling firm
Med-Hot Thermal Imaging, Inc.
Product
MEDHOT MTI 2000 THERMAL IMAGING SYSTEM - TotalVision software with Elevated Body Temperature Screening module, for use as a medical diagnostic purposes as part of a telethermographic system utilizing FLIR cameras, Models A615sc and A315sc. Cameras are private labeled as MAX Basic, MAX 76 and MAX 307.
Status
Terminated
Initiated
2017-01-14
Posted
not on file
Terminated
2019-08-07
Reason
Indications for use are not cleared or approved for market in the U.S.
Root cause
Labeling Change Control
Action
Correct software to eliminate any reference to temperature. Test function and notify users of software update with new user instructions.
Quantity
12
Distribution
US: AL and NJ - Foreign: Nigeria
Lot, serial or model codes
P/N 18001-EBT-MP-D with FLIR A315 camera P/N 15102-EBT-MP-D wjth FLIR A615 camera TotalVision Software versions 1.6 1.7.1
510(k) numbers
["K063047"]
PMA numbers
not on file
Product code
LHQ
Firm FEI
3006033213
Firm city
Lakeland
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register