Accuray Inc: CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software versions 1.5.2 and higher.
- Recall number
- Z-2056-2008
- Recalling firm
- Accuray Inc
- Product
- CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software versions 1.5.2 and higher.
- Status
- Terminated
- Initiated
- 2008-04-09
- Posted
- 2008-09-17
- Terminated
- 2010-11-30
- Reason
- Sample beam data (which should not be used to treat patients) may differ from actual radiation output of an installed product, and which may be used by users.
- Root cause
- Software design
- Action
- On April 16, 2008, the firm issued notifications via fedex to affected sites. In addition, installation and upgrade procedures will be updated to incorporate a step which decommissions sample beam data. If you have questions, contact Accuray Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or send and email to customersupport@accuray.com.
- Quantity
- 109 units
- Distribution
- Product was distributed to 104 medical centers throughout the US and to Vietnam, China, Korea, Taiwan, Malaysia, Turkey, France, Italy, Spain, Greece and Germany.
- Lot, serial or model codes
- Versions 1.5.2, with system numbers (all preceded by C0) 095, 142, 058, 106, 061, 056, 128, 129, 079, 088, 116, 151, 139, 047, 070, 072, 133, 112, 149, 078, 137, 025, 098, 092, 045, 144, 124, 160, 044, 126, 123, 111, 055, 062, 074, 068, 059, 052, 103, 132, 100, 150, 051, 094, 119, 034, 146, 153, 064, 155, 037, 114, 046, 063, 097, 143, 067, 131, 042, 048, 090, 091, 122, 159, 022, 109, 099, 135, 036, 134, 127, 158, 057, 069, 030, 118, 108, 152, 136, 049, 102, 080, 101, 141, 071, 087, 140, 107, 145, 148, 120, 023, 147, 089, 038, 041, 065, 138, 060, 113, 110, 130, 043, 125, 105, 093, 117, 032, 066
- 510(k) numbers
- ["K072504"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1000123585
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28