Skip to the content

FDA Medical Device Recalls (openFDA)

Accuray Inc: CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software versions 1.5.2 and higher.

Recall number
Z-2056-2008
Recalling firm
Accuray Inc
Product
CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software versions 1.5.2 and higher.
Status
Terminated
Initiated
2008-04-09
Posted
2008-09-17
Terminated
2010-11-30
Reason
Sample beam data (which should not be used to treat patients) may differ from actual radiation output of an installed product, and which may be used by users.
Root cause
Software design
Action
On April 16, 2008, the firm issued notifications via fedex to affected sites. In addition, installation and upgrade procedures will be updated to incorporate a step which decommissions sample beam data. If you have questions, contact Accuray Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or send and email to customersupport@accuray.com.
Quantity
109 units
Distribution
Product was distributed to 104 medical centers throughout the US and to Vietnam, China, Korea, Taiwan, Malaysia, Turkey, France, Italy, Spain, Greece and Germany.
Lot, serial or model codes
Versions 1.5.2, with system numbers (all preceded by C0) 095, 142, 058, 106, 061, 056, 128, 129, 079, 088, 116, 151, 139, 047, 070, 072, 133, 112, 149, 078, 137, 025, 098, 092, 045, 144, 124, 160, 044, 126, 123, 111, 055, 062, 074, 068, 059, 052, 103, 132, 100, 150, 051, 094, 119, 034, 146, 153, 064, 155, 037, 114, 046, 063, 097, 143, 067, 131, 042, 048, 090, 091, 122, 159, 022, 109, 099, 135, 036, 134, 127, 158, 057, 069, 030, 118, 108, 152, 136, 049, 102, 080, 101, 141, 071, 087, 140, 107, 145, 148, 120, 023, 147, 089, 038, 041, 065, 138, 060, 113, 110, 130, 043, 125, 105, 093, 117, 032, 066
510(k) numbers
["K072504"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1000123585
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register