FDA Medical Device Recalls (openFDA)
Smith & Nephew Inc: R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 52 MM O.D., REF 71338952, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
- Recall number
- Z-2055-2011
- Recalling firm
- Smith & Nephew Inc
- Product
- R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 52 MM O.D., REF 71338952, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
- Status
- Terminated
- Initiated
- 2011-03-11
- Posted
- 2011-04-22
- Terminated
- 2012-02-08
- Reason
- During the manufacturing process for several batches of R3 Ceramic Liners, the titanium rings were pressed onto the ceramic component with a higher force than allowed by manufacturing specifications. This has the potential to result in lower than expected strength for the liners.
- Root cause
- Employee error
- Action
- All affected Smith & Nephew Distributors were notified via email and the hospitals were notified by letter delivered by Fed-Ex on 3/11/2011. They were instructed to immediately cease distribution, quarantine affected product for return, if product further distributed, identify customers and immediately notify them of recall. Please contact Mr. David Archer, Group Director, at 701.396.2121, ext. 5373, if you have questions.
- Quantity
- 563 units
- Distribution
- AK, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SC, TN, TX, WA, and WI
- Lot, serial or model codes
- 09FT32383, 09FT32504, 09GT33518, 09GT33519, 09GT33918, 09GT33919, 09GT34111, 09GT34112, 09KT36378, and 09MT38082
- 510(k) numbers
- not on file
- PMA numbers
- ["P030022"]
- Product code
- MRA
- Firm FEI
- 1020279
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as