Hitachi Medical Systems America Inc: EUB-9500/HI VISION 8500 Diagnostic Ultrasound Scanners, DICOM Option EZU-FC6, Software Versions: V11-01, V11-03, V12-03 and V12-04A.

Recall number
Z-2055-2009
Recalling firm
Hitachi Medical Systems America Inc
Product
EUB-9500/HI VISION 8500 Diagnostic Ultrasound Scanners, DICOM Option EZU-FC6, Software Versions: V11-01, V11-03, V12-03 and V12-04A.
Status
Terminated
Initiated
2008-05-12
Posted
2009-09-09
Terminated
2011-07-20
Reason
The firm's parent company discovered a DICOM UID (Universal Identifier) code mix-up which can occur due to a software error. This error can cause the images from two (2) different patient scans to become intermingled and consequently, labeled incorrectly.
Root cause
Software design
Action
HMSA issued a 'DEVICE CORRECTION' notification to their customers on 5/12/2008. The notification describes the potential data transmission problem (potential to archive studies done on different patients into the same patient folder) and the conditions under which this problem can occur. The notification cautions the user/customer to verify all images stored on the PACS (Picture Archiving and Communications System) in order to ensure that they do not contain images from a different patient. The DEVICE CORRECTION letter also states that the firm will be installing updated software on all affected systems beginning in June, 2008, in order to address this issue. It also states that a Hitachi Service Rep. will be contacting each customer and making arrangements to set up a convenient time to come out and perform a software upgrade. Lastly, the letter provides a toll-free telephone # to the firm's 'Customer Service Center: 1-800-800-4925, which the customer can use to obtain additional information, if desired.
Quantity
6 units
Distribution
The recalled device system units were distributed to customers located in the following states within the U.S ony.: AR, CA, and TX.
Lot, serial or model codes
The following Serial Numbers of this device were subject to recall/software correction: 17084306, 17084308, 19921303, 17268307, 17268309, and 18190406.
510(k) numbers
["K013722"]
PMA numbers
not on file
Product code
ITX
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register