Siemens Healthcare Diagnostics, Inc.: ADVIA Centaur TSH3-Ultra (Thyroid Stimulating Hormone 3 Ultra), Catalog Numbers: 06491072 (100 tests), 06491080 (500 tests), 04862625 (2500 tests)

Recall number
Z-2054-2018
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
ADVIA Centaur TSH3-Ultra (Thyroid Stimulating Hormone 3 Ultra), Catalog Numbers: 06491072 (100 tests), 06491080 (500 tests), 04862625 (2500 tests)
Status
Open, Classified
Initiated
2017-12-18
Posted
not on file
Terminated
not on file
Reason
In certain scenarios the ADVIA Centaur CP, XP and XPT immunoassay analyzers may exhibit reduced dilution recovery or become inoperable when used with Multi-Diluent 15 and certain lots of BNP and TSH3RUL Assays.
Root cause
Device Design
Action
Consignees are asked to return the response form and follow the instructions pertaining to their system and assays: Actions for ALL ADVIA Centaur Systems: Do not perform onboard dilutions on any ADVIA Centaur system if any of the affected kits are present on the system. Onboard dilutions for the ADVIA Centaur BNP and/or TSH3UL assay(s) are only supported when using ADVIA Centaur BNP kit lots ending in 210 and above and ADVIA Centaur TSH3UL kit lots ending in 310 and above. Actions for ADVIA Centaur XPT system: If you intend to perform onboard dilutions for the ADVIA Centaur BNP and/or TSH3UL assays, you must restart the workstation after scanning the master curve card for the BNP and/or TSH3UL assay kit lot, before ordering an onboard dilution for these assays. Please perform the attached steps to restart the workstation. Actions for ADVIA Centaur CP system: If any of the affected kit lots have been in use on the system, you must manually track the time that the Multi-Diluent 15 has been in use on board the system. Replace the ancillary pack of Multi-Diluent 15 every 7 days when used, until a follow-up communication is received from Siemens stating this issue has been resolved.
Quantity
74,215
Distribution
Distributed in the following U.S. states: AL, CA, DE, GA, IA, IL, IN, KY, MA, MI, MS, MT, NC, ND, NJ, NM, NV, NY, OH, PA, RI, TN, TX, VA, WA, WV, WY and Puerto Rico. Distributed in the following OUS countries: Australia Belgium¿¿¿¿¿¿¿¿ Canada Chile France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Israel Italy¿¿¿¿¿¿¿¿¿¿ Japan Latvia¿¿¿¿¿¿¿¿¿ Luxembourg¿¿¿¿¿ Netherlands¿¿¿¿ New Zealand Norway¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Singapore Slovakia¿¿¿¿¿¿¿ South Africa Spain¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ U.A.E.¿¿¿¿¿¿¿¿¿ United¿Kingdom¿
Lot, serial or model codes
Lot Numbers ending in: 298 and 301
510(k) numbers
["K150403"]
PMA numbers
not on file
Product code
JLW
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register