Philips North America, LLC: Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.
- Recall number
- Z-2053-2020
- Recalling firm
- Philips North America, LLC
- Product
- Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.
- Status
- Terminated
- Initiated
- 2020-04-13
- Posted
- not on file
- Terminated
- 2021-04-07
- Reason
- The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with an energy level different than the setting selected by the user. In any of these scenarios, appropriate therapy delivery may be delayed.
- Root cause
- Process control
- Action
- On April 13, 2020, the firm distributed URGENT Medical Device Recall letters to affected customers. Customers were informed of the device defect, which may result in the following unexpected behavior: - The device may not turn on. - The device may not perform the selected function. - The device may deliver a shock with an energy level different from the setting selected by the user. There have been three reported patient deaths potentially associated with the failure of the HeartStart XL Monitor/Defibrillator rotary therapy selector switch. Customers were advised that the HeartStart XL Defibrillator/Monitor has been discontinued and has reached its end of life. Therefore, custoemrs should replace and retire their units as soon as practically possible. Customers should continue to perform Shift Checks and Operational checks as recommended in the Instructions for Use (IFU) as this reduces the risk of a failure until your units are safely retired. If you identify a device that exhibits any of the behaviors described above, or which fails shift or operational checks, immediately remove it from service. Customers were asked to complete and fax or email the included customer reply form to Philips. If you need further information or support concerning this notification, please contact your local Philips representative or call us at 1-800-722-9377.
- Quantity
- 94,034 units
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Lot, serial or model codes
- All units. Product has been discontinued.
- 510(k) numbers
- ["K001725"]
- PMA numbers
- not on file
- Product code
- MKJ
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28