Leica Microsystems, Inc.: Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage: The intended use of this device is for improving visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

Recall number
Z-2053-2012
Recalling firm
Leica Microsystems, Inc.
Product
Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage: The intended use of this device is for improving visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
Status
Terminated
Initiated
2012-06-04
Posted
2012-07-20
Terminated
2012-10-29
Reason
Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a remote video adapter or motorized dual video adapter.
Root cause
Software design
Action
Leica Microsystems sent an Urgent Medical Device letter dated June 4, 2012 to the affected customers. The letter described the product, problem and actions to be taken. The customers were informed that the Leica surgical microscope M525 F50 required a software upgrade. Customers were instructed to NOT USE a remote video adapter or motorized video adapter until the required software upgrade has been installed by their Leica Microsystems representative. The customers were requested to complete the attached Acknowledgement Form, acknowledging receipt and understanding of the letter, listing the serial numbers of the units at their facilities and to fax it back to Leica Microsystems at (847) 236-3747.
Quantity
39 Units - USA , 4 Units - Canada
Distribution
Worldwide Distribution - US (nationwide) in the states of: California, Colorado, Florida, Georgia, Illinois, Indiana, Massachusetts, Michigan, Minnesota, New York, Oklahoma, Oregon, Pennsylvania, Texas, Utah, West Virginia, including Puerto Rico, and the countriy of Canada.
Lot, serial or model codes
Serial numbers serial numbers 110311001, 120312001, 20412001, 260112001, 300911001, 40811001, 80411001, 90312001, 10311002, 90611001, 260711001, 20911001, 200911001, 210911001, 111011001, 141011001, 241011001, 121211001, 141211001, 151211001, 191211001, 30112001, 50112001, 140212001, 60312001, 220312001, 300312001, 30412001, 50411001, 10411001, 130411001, 210411001, 170511001, 300511001, 90811001, 200312001, 50412001, 110112002, 280611001, 241011002, 20312001, 101011001 and 270911001
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSO
Firm FEI
3000210133
Firm city
Buffalo Grove
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register