Skip to the content

FDA Medical Device Recalls (openFDA)

Integra Neuro Sciences: Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit NL950-V and Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit with Ventricular Drainage System NL950-VC. This intracranial pressure (ICP) catheter is indicated for use on patients that require continuous invasive intracranial pressure monitoring and/or cerebrospinal drainage. Access is achieved through a tunneled technique using the included trocar.

Recall number
Z-2053-2009
Recalling firm
Integra Neuro Sciences
Product
Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit NL950-V and Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit with Ventricular Drainage System NL950-VC. This intracranial pressure (ICP) catheter is indicated for use on patients that require continuous invasive intracranial pressure monitoring and/or cerebrospinal drainage. Access is achieved through a tunneled technique using the included trocar.
Status
Terminated
Initiated
2009-03-16
Posted
2009-09-08
Terminated
2012-07-19
Reason
Due to a dimensional error, the diameter of the stylet provided with the device may be too large to insert into the lumen of the catheter tip.
Root cause
Process design
Action
An "Urgent-Product Recall Notification" letter dated March 23, 2009 was issued to consignees. The letter described the issue and instructed customers to examine their inventory immediately to determine whether they have any NL950V - Ventrix Ventricular Tunneling Pressure Monitoring Kits, or NL950VC - Ventrix Ventricular Tunneling Pressure Monitoring Kits with Ventricular Drainage System. Customers were further instructed not to use the above recalled product. Call Integra Customer Service at 1-800-654-2873, and then select 5 for 'Returns and Repairs' to obtain a Returned Material Authorization number, and return the product as directed. Direct questions about the recall to Integra NeuroSciences by calling 1-858-455-1115, extension185.
Quantity
274 kits
Distribution
Worldwide Distribution -- US (states of AR, CA, OH, TX, CT, NY, LA, FL, NY, IL, and VA), Belgium and Malaysia.
Lot, serial or model codes
NL950-V - Lot Numbers: 30500R130474 (exp. 8/31/2011); 30500R134098 (exp. 10/31/2011); 30500R139222 (exp. 11/30/2011); 30500R143588 (exp. 1/31/2012); 30500R146383 (exp. 1/31/2012); 30500R147342 (exp. 2/29/2012); and 30500R147899 (exp. 2/29/2012) and  NL950-VC - Lot Numbers: 138999 (exp. 9/30/2011); and 142754 (exp. 12/31/2011).
510(k) numbers
["K904883"]
PMA numbers
not on file
Product code
JXG
Firm FEI
2023988
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register