FDA Medical Device Recalls (openFDA)
Thoratec Corporation: HeartMate II LVAS with Pocket Controller; 106015 - HMII LVAD SEALED KIT-PCTRL, NA UDI: 00813024011224 Intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; intended for use inside or outside the hospital.
- Recall number
- Z-2052-2017
- Recalling firm
- Thoratec Corporation
- Product
- HeartMate II LVAS with Pocket Controller; 106015 - HMII LVAD SEALED KIT-PCTRL, NA UDI: 00813024011224 Intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; intended for use inside or outside the hospital.
- Status
- Terminated
- Initiated
- 2017-03-30
- Posted
- 2017-05-24
- Terminated
- 2019-03-20
- Reason
- St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
- Root cause
- Device Design
- Action
- Urgent Medical Device Correction letters, dated March 29, 2017, were sent to customers. The letters provided information on what the current correction will involve to ensure patients are successfully able to exchange their pocket controller in an emergency situation. Patients will receive the updates at scheduled appointment. Questions regarding the notice, should be directed to Abbott HeartLine at 800-456-1477; or, your Abbott Sales Representative.
- Quantity
- 24,077 active units in US
- Distribution
- Worldwide Distribution -- US and Austria, Belgium, Brazil, Canada, Cayman Islands, Chile, Columbia, Croatia, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Kazakhstan, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Malaysia, Orman, Palastine, Poland, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Thailand, The Netherlands, Turkey, United Arab Emirates, and United Kingdom.
- Lot, serial or model codes
- Serial numbers PC-14810, PC-14828, PC-14843, PC-14846, and PC-14848.
- 510(k) numbers
- not on file
- PMA numbers
- ["P060040"]
- Product code
- DSQ
- Firm FEI
- 2916596
- Firm city
- Pleasanton
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28