Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc.: ACUSON S2000 ultrasound system Manufactured by Siemens Medical Solutions USA, Inc. Business Unit Ultrasound, 1230 Shorebird Way Intended for the following applications: Fetal, abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal and Peripheral Vascular Applications.

Recall number
Z-2051-2011
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ACUSON S2000 ultrasound system Manufactured by Siemens Medical Solutions USA, Inc. Business Unit Ultrasound, 1230 Shorebird Way Intended for the following applications: Fetal, abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal and Peripheral Vascular Applications.
Status
Terminated
Initiated
2011-03-15
Posted
2011-04-22
Terminated
2012-06-18
Reason
The Firm discovered a software error on the Accuson S2000 ultrasound systems affecting the measurement results in the patient report.
Root cause
Software design
Action
Siemens issued a Customer Safety Advisory Notification letter to all customers on March 15, 2011, regarding a software error on the ACUSON S2000 ultrasound systems affected measurement . They are in the process of releasing a software update that will correct this issue. Siemens recommends reviewing the patient report at the end of each study to confirm that the report contains only measurements relevant to the exam just performed. No patient injury has been reported. Further questions regarding this recall please call (650) 694-5398.
Quantity
1533 units
Distribution
Worldwide Distribution including USA, states of AZ, VA, CA, NE, IL, MN, TN, FL, NM, and DC
Lot, serial or model codes
Model Number 10041461-ACUSON S2000. Serial numbers to follow
510(k) numbers
["K082142"]
PMA numbers
not on file
Product code
IYN
Firm FEI
2936884
Firm city
Mountain View
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register