Linvatec Corp. dba ConMed Linvatec: REF IA-2000-S, LightWave Suction Ablator, STERILE, Rx ONLY, MADE IN USA. CONMED LINVATEC, 11311 CONCEPT BLVD, LARGO, FL 33773-4908. ConMed Linvatec LIGHTWAVE SUCTION ABLATOR, electrosurgical cutting and coagulation device and accessories, intended to remove tissue and control bleeding by use of high-frequency electrical current.

Recall number
Z-2051-2009
Recalling firm
Linvatec Corp. dba ConMed Linvatec
Product
REF IA-2000-S, LightWave Suction Ablator, STERILE, Rx ONLY, MADE IN USA. CONMED LINVATEC, 11311 CONCEPT BLVD, LARGO, FL 33773-4908. ConMed Linvatec LIGHTWAVE SUCTION ABLATOR, electrosurgical cutting and coagulation device and accessories, intended to remove tissue and control bleeding by use of high-frequency electrical current.
Status
Terminated
Initiated
2009-07-31
Posted
2009-09-09
Terminated
2011-02-25
Reason
ConMed Linvatec has determined there is a possibility that the device may not shut off if the ablate button is released, which may result in users/patients suffering burns.
Root cause
Device Design
Action
ConMed Linvatec is notifying their direct consignees by an Urgent Medical Device Recall letter, requesting that customers check their inventory for the affected products; complete, sign, and return the Reply Form; and return of any affected lot numbers of the product. The product is to be returned to ConMed Linvatec. Questions should be directed to Customer Service at 1-800-237-0169.
Quantity
28,074 units
Distribution
Worldwide Distribution
Lot, serial or model codes
All lots manufactured from July 2004 to September 2008: Lot Numbers: 0407121, 0407151, 0407211, 0407271, 0411151, 0411221, 0411291, 0412081, 0412111, 0502161, 0502211, 0504281, 0505021, 0505061, 0505091, 0505101, 0505121, 0505161, 0505171, 0505231, 0506021, 0506131, 0507051, 0508121, 0508221, 0508251, 0509061, 0511021, 0511091, 0511301, 0602231, 0604241, 0606071, 0606191, 0610271, 0611011, 0703121, 0703191, 0704021, 0704091, 0704121, 0705091, 0705231, 0706041, 0706221, 0706271, 0707241, 0708031, 0708131, 0708201, 0711261, 0711301, 0801072, 0801152, 0802132, 0802202, 0803072, 0803112, 0803262, 0803312, 0804171, 0804292, 0805022, 0805061, 0805062, 0806092, 0806262, 0807142, 0807232, 0807292, 0808052, 0808132, 0809032, 0809052, and 0809262.
510(k) numbers
["K050923"]
PMA numbers
not on file
Product code
GEI
Firm FEI
1017294
Firm city
Largo
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register