FDA Medical Device Recalls (openFDA)
Roche Diagnostics Corporation: Sample probes associated with: cobas 6000 Series System, cobas 8000 Modular Analyzer Series, cobas c 311 analyzer cobas c 501/502 cobas c 701/702 cobas 8000 ISE Product Usage: The cobas 6000 series is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of Ion selective electrodes (ISE), a photometric analysis unit an immunoassay analysis module. The cobas 8000 Modular Analyzer Series is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit. The cobas c 311 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
- Recall number
- Z-2049-2017
- Recalling firm
- Roche Diagnostics Corporation
- Product
- Sample probes associated with: cobas 6000 Series System, cobas 8000 Modular Analyzer Series, cobas c 311 analyzer cobas c 501/502 cobas c 701/702 cobas 8000 ISE Product Usage: The cobas 6000 series is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of Ion selective electrodes (ISE), a photometric analysis unit an immunoassay analysis module. The cobas 8000 Modular Analyzer Series is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit. The cobas c 311 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
- Status
- Terminated
- Initiated
- 2017-03-21
- Posted
- not on file
- Terminated
- 2017-10-02
- Reason
- Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the affected sample probe may dip into the sample material deeper than intended. This may lead to carryover and medical risk cannot be excluded.
- Root cause
- Component change control
- Action
- On March 21, 2017, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.
- Quantity
- 3946
- Distribution
- US Nationwide.Distribution
- Lot, serial or model codes
- sample probe catalog numbers: 04547241001 - cobas c 501/502 05899427001 - cobas c 311 04945794001 - cobas c 701/702 and cobas 8000 ISE Lot numbers: cobas c 311 0490232027, 0490233855, 0490236427, 0490238543, 0490241937, 0490248322, 0490250880, 0490235713, 0490252029, 0490256963 cobas c 501, c 502 0490233762, 0490233764, 0490233766, 0490233767, 0490233777, 0490235904, 0490235905, 0490235906, 0490235907, 0490235949, 0490235950, 0490237594, 0490237595, 0490237353, 0490237354, 0490237597, 0490237598, 0490237600, 0490237602, 0490237604, 0490237606, 0490238955, 0490238957, 0490238893, 0490238895, 0490238896, 0490238898, 0490238899, 0490242149, 0490242151, 0490242153, 0490242155, 0490242327, 0490242328, 0490242329, 0490242330, 0490242331, 0490250491, 0490250492, 0490250493, 0490250494, 0490250495, 0490250496, 0490251032, 0490251033, 0490251034, 0490251036, 0490252524, 0490252527, 0490252529, 0490252532, 0490252247, 0490252249, 0490252251, 0490252252, 0490252272, 0490252398, 0490254550, 0490254551, 0490254552, 0490254553, 0490254554, 0490254555, 0490254804, 0490254805, 0490254806, 0490254807, 0490256999, 0490257000, 0490256993, 0490256994, 0490256995, 0490256996, 0490256997, 0490256998, 0490254808, 0490257001, 0490257847, 0490257848, 0490257849, 0490257850, 0490257851, 0490258652 cobas c 701, c 702 0490238960, 0490238961, 0490238963, 0490242332, 0490251040, 0490251041, 0490252546, 0490252547, 0490254810, 0490254811, 0490221658, 0490223228, 0490223229, 0490225255, 0490226172, 0490226173, 0490226174, 0490227626, 0490227627, 0490227628, 0490230276, 0490230277, 0490230279, 0490232203, 0490232204, 0490232205, 0490232206, 0490233778, 0490233779, 0490233780, 0490233781, 0490233783, 0490235909, 0490235910, 0490235911, 0490235912, 0490235913, 0490237365, 0490237366, 0490237367, 0490237368, 0490237369, 0490237370, 0490238901, 0490238902, 0490238903, 0490238904, 0490242160, 0490242161, 0490242163, 0490242165, 0490242166, 0490250498, 0490250499, 0490250500, 0490250501, 0490250502, 0490252308, 0490252312, 0490252318, 0490252320, 0490252322, 0490254558, 0490254559, 0490254560, 0490254561, 0490254562, 0490257035, 0490257036, 0490257037, 0490257038, 0490257039, 0490257854, 0490257855, 0490257856, 0490257857, 0490257858
- 510(k) numbers
- ["K060373","K100853"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28