Skip to the content

FDA Medical Device Recalls (openFDA)

Nidek Inc: NAVIS-EX , Image Filing Software for NIDEK AFC Fundus camera, Ver 1.1.0 to 1.3.5. NAVIS-EX is indicated for use as file management software for ophthalmic images.

Recall number
Z-2046-2013
Recalling firm
Nidek Inc
Product
NAVIS-EX , Image Filing Software for NIDEK AFC Fundus camera, Ver 1.1.0 to 1.3.5. NAVIS-EX is indicated for use as file management software for ophthalmic images.
Status
Terminated
Initiated
2013-08-01
Posted
2013-08-23
Terminated
2014-02-25
Reason
Under certain circumstances, the Left eye image may be saved as the Right eye image.
Root cause
Software design
Action
"Medical Device Correction Notification" letters dated August 5, 2013 were sent to consignees through their US contracted distributor - Marco Opthalmic. The letter described the product issue and provided information on the correction. Customers can contact 800-223-9044 for questions.
Quantity
86
Distribution
Nationwide US distribution only.
Lot, serial or model codes
version 1.1.0; serial numbers: 320823 320425 320622 121424 320094 320606 320617 320577 320636 320630 320468 320489 320518 320695 320695 320652 320666 320692 320675 320640 320492 320695 320704 320777 320779 320771 320768 320753 320744 320780 320617; version 1.2.1; serial numbers 320740 120045; Version 1.3.0: 141356 320814 320815 140941 141326 320847 320916 320906 320203 320569 320809 320813 141357 320831 140937 141358 320798 320090 320980 320990 320927 320942 320920 141401 320983; Version 1.3.4; 321035 321020 320998 141453; Version 1.3.5: 380215 380188 380189 380251 380211 380014 380077 380213 380076 380034 380035 380017 380244 380124 380243 380075 380122 380021 380253/380052 380221/380214 141480 141471 321053 320883.
510(k) numbers
["K113451"]
PMA numbers
not on file
Product code
HKI
Firm FEI
2936921
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register