FDA Medical Device Recalls (openFDA)
Fujifilm Medical Systems U.S.A., Inc.: Fujifilm Synapse PACS software version 4.4.000, Fujifilm Synapse PACS software version 4.4.001, Fujifilm Synapse PACS software version 4.4.004, Fujifilm Synapse PACS software version 4.4.010 and Fujifilm Synapse PACS software version 4.4.020 FUJIFILM Synapse Workstation Software (a Picture Archiving And Communications System) is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. Also intended for installation on an off-the-shelf PC networked with Fuji Synapse PACS.
- Recall number
- Z-2043-2016
- Recalling firm
- Fujifilm Medical Systems U.S.A., Inc.
- Product
- Fujifilm Synapse PACS software version 4.4.000, Fujifilm Synapse PACS software version 4.4.001, Fujifilm Synapse PACS software version 4.4.004, Fujifilm Synapse PACS software version 4.4.010 and Fujifilm Synapse PACS software version 4.4.020 FUJIFILM Synapse Workstation Software (a Picture Archiving And Communications System) is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. Also intended for installation on an off-the-shelf PC networked with Fuji Synapse PACS.
- Status
- Terminated
- Initiated
- 2016-05-10
- Posted
- not on file
- Terminated
- 2019-02-20
- Reason
- Synapse cannot display image files, DICOM SR files, and/or Annotation files. The "Image Not Loaded" message displays instead
- Root cause
- Software design
- Action
- The firm, FujiFilm, sent an "MEDICAL DEVICE CORRECTION" letter dated May 10, 2016 via Federal Express to its customers. The letter describes the product, problem and actions to be taken.The customers were instructed to: Extend the Ad hoc deletion retention period or discontinue use of Ad Hoc Deletion, and complete and submit the attached Customer Response Form to Fujifilm via fax to (203)-251-7863. Affected systems will be corrected via on-site visits by FMSU Customer Service Engineers upgrading to Version 4.4.100. Questions or concerns, contact the FUJI FILM Customer Support Center at 1- 888-FUJIMED Option #1 .
- Quantity
- 27 units
- Distribution
- US Distribution to states of: CA, FL, KS, NE, NY, and PA.
- Lot, serial or model codes
- Software version: 4.4.000, 4.4.001, 4.4.004, 4.4.010, 4.4.020
- 510(k) numbers
- ["K112439"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 1000513161
- Firm city
- Stamford
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28