Skip to the content

FDA Medical Device Recalls (openFDA)

Immuno-Mycologics, Inc: Histoplasma Yeast Complement Fixation Antigen Dilute Ref# H30150 The fungal antigens and positive controls are used to detect antibodies in patient serum by the complement fixation (CF) procedure to aid in the diagnosis of four specific fungal diseases: Histoplasmosis, blastomycosis, coccidioidomycosis, and aspergillosis

Recall number
Z-2040-2016
Recalling firm
Immuno-Mycologics, Inc
Product
Histoplasma Yeast Complement Fixation Antigen Dilute Ref# H30150 The fungal antigens and positive controls are used to detect antibodies in patient serum by the complement fixation (CF) procedure to aid in the diagnosis of four specific fungal diseases: Histoplasmosis, blastomycosis, coccidioidomycosis, and aspergillosis
Status
Terminated
Initiated
2016-05-09
Posted
2016-06-20
Terminated
2017-01-10
Reason
The device was found to have bacterial contamination.
Root cause
Process control
Action
Immuno-Mycologics, Inc sent an Urgent Medical Device Recall letter dated May 9, 2016, to all affected customers. The letter identified the affected product, the problem, and the action to be taken by the customer. Customers were instructed to immediately check their stock and destroy affected stock on hand. Immediately complete the attached Acknowledgement and Receipt Form even if they do not have affected stock. Customers who require replacement indicate the number of vials needed in the attached Acknowledgement Form. Customers with questions were instructed to call 1-405-360-4669.
Quantity
467 units
Distribution
Nationwide Distribution to UT, MA, OH, IN, NC, NY, VA, CA, IA and KY.
Lot, serial or model codes
Lot 111WH3
510(k) numbers
["K791393"]
PMA numbers
not on file
Product code
GMJ
Firm FEI
1627497
Firm city
Norman
Firm state
OK
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register