Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Vascular: NC Traveler RX Coronary Dilatation Catheter, Part No. 1013152-12, 1013152-15, 1013152-20, 1013153-08, 1013153-12, 1013153-15, 1013153-20, 1013154-08, 1013154-12, 1013154-15, 1013154-20, 1013155-08, 1013155-12 1013155-15, 1013155-20, 1013156-12, 1013156-15, 1013157-08, 1013157-12, 1013157-15

Recall number
Z-2038-2017
Recalling firm
Abbott Vascular
Product
NC Traveler RX Coronary Dilatation Catheter, Part No. 1013152-12, 1013152-15, 1013152-20, 1013153-08, 1013153-12, 1013153-15, 1013153-20, 1013154-08, 1013154-12, 1013154-15, 1013154-20, 1013155-08, 1013155-12 1013155-15, 1013155-20, 1013156-12, 1013156-15, 1013157-08, 1013157-12, 1013157-15
Status
Terminated
Initiated
2017-03-22
Posted
2017-05-12
Terminated
2018-02-07
Reason
Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon.
Root cause
Device Design
Action
An Urgent Field Safety Notice dated 3/22/17 was sent to customers to inform them that Abbott Vascular is recalling the NC Trek RX Coronary Dilation Catheter, NC Traveler RX Coronary Dilatation Catheter, and NC Tenku RX PCTA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon. The letter informs the customers of how does the issue occur and what action is Abbott Vascular asking the customers to take. Customers with any questions are instructed to contact local Abbott Vascular Representative or Customer Service Department at (800) 227-9902. On 5/15/17, Abbott Vascular posted a press release on their website to inform customers that Abbott has initiated a voluntary recall of specific lots of three catheters: NC Trek RX Coronary Dilatation Catheter, NC Traveler Coronary Dilatation Catheter, and NC Tenku RX PTCA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath, which could cause problems with inflating or deflating the balloon. Potential risks associated with balloon inflation and deflation difficulties include air embolism, additional intervention, thrombosis, and myocardial infarction.
Quantity
449,661 units total (132,040 units in US)
Distribution
US and Worldwide
Lot, serial or model codes
Lot Number 61209G1 61209G2 61102G1 61102G1 61012G1 61209G1 61101G1 60415G1 61209G1 61102G1 61101G1 61101G1 61102G1 60415G1 61102G1 61101G1 60415G1 61209G1 61102G1 50415G1 50919G1 51110G1 61209G1 50919G1 51210G1 61101G1 61102G1 60415G1 61102G1 60415G1 61209G1
510(k) numbers
["K103153"]
PMA numbers
not on file
Product code
LOX
Firm FEI
2024168
Firm city
Temecula
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register