Skip to the content

FDA Medical Device Recalls (openFDA)

Hitachi Medical Systems America Inc: Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Recall number
Z-2037-2012
Recalling firm
Hitachi Medical Systems America Inc
Product
Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Status
Terminated
Initiated
2012-05-04
Posted
2012-07-19
Terminated
2013-02-19
Reason
Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.
Root cause
Software design
Action
Hitachi sent a "Device Correction Notice" dated May 4, 2012 to their customers. The notice identified the affected product, problem and actions to be taken. Customers were instructed to "not use the Reconstruction Status" function to suspend active reconstructions until the software problem can be corrected. They also informed their customers that a representative will visit their facility to install the software update to permanently resolve the issue at no cost to the facility. For further information regarding this notice call 800-800-3106 x3720.
Quantity
Six (6) units
Distribution
US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX.
Lot, serial or model codes
Serial #'s S5001 - S5011
510(k) numbers
["K101888"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register