Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.

Recall number
Z-2036-2012
Recalling firm
Teleflex Medical
Product
Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.
Status
Terminated
Initiated
2012-06-12
Posted
2012-07-19
Terminated
2014-01-28
Reason
Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.
Root cause
Device Design
Action
Teleflex sent Urgent Medical Device Notification letters dated June 12, 2012 to all affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to complete the enclosed Acknowledgement Form and return per the instructions provided. Once the form is received, Teleflex Medial will schedule for a technician to come to their facility and exchange the affected product. For questions call your local sales representative or Customer Service at 1-866-246-6990.
Quantity
14,794 each
Distribution
Worldwide Distribution - U.S. (nationwide) and the countries of Argentina, Australia, Brazil, Canada, Chile, China, Europe, India, Japan, Mexico, South Korea, Taiwan and UK.
Lot, serial or model codes
Catalog #425-00, 425-10 and 425-30.
510(k) numbers
["K063758"]
PMA numbers
not on file
Product code
BTT
Firm FEI
3005747797
Firm city
Research Triangle Park
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register