Skip to the content

FDA Medical Device Recalls (openFDA)

The Anspach Effort, Inc.: ***REF MCA5-2SB*** 2 mm Fluted Ball, 5.6 cm. Use with Micro Curved Attachment (MCA).***ANSPACH***Sterile, Sterilized using Irradiation***Rx Only*** Made in the USA***Manufacturer: The Anspach Effort, Inc. 4500 Riverside Drive, Palm Beach Gardens, FL 33410 USA Tele: (800) 327-6887/ + 1 561 627 1080***Anspach Europe Ltd. Aston Court, Kingsmead Business Park Frederick Place High Wycombe, HP11 1LA United Kingdom Tel: +441494 616126*** Lot numbers: C523029666, D063031864, D113032937, D113032974, D163033908, D253036524, and D353039102. Intended usage: Cutting and shaping bone including bones of the spine and cranium.

Recall number
Z-2036-2011
Recalling firm
The Anspach Effort, Inc.
Product
***REF MCA5-2SB*** 2 mm Fluted Ball, 5.6 cm. Use with Micro Curved Attachment (MCA).***ANSPACH***Sterile, Sterilized using Irradiation***Rx Only*** Made in the USA***Manufacturer: The Anspach Effort, Inc. 4500 Riverside Drive, Palm Beach Gardens, FL 33410 USA Tele: (800) 327-6887/ + 1 561 627 1080***Anspach Europe Ltd. Aston Court, Kingsmead Business Park Frederick Place High Wycombe, HP11 1LA United Kingdom Tel: +441494 616126*** Lot numbers: C523029666, D063031864, D113032937, D113032974, D163033908, D253036524, and D353039102. Intended usage: Cutting and shaping bone including bones of the spine and cranium.
Status
Terminated
Initiated
2010-09-29
Posted
2011-04-20
Terminated
2011-08-12
Reason
Anspach Effort, Inc. Palm Beach Gardens, FL is recalling their Sterile Bone Cutting Burs that was manufactured with incomplete flutes which could cause degraded cutting performance and in some cases not cut at all. Anspach has not received any complaints related to this problem.
Root cause
Other
Action
The firm, Anspach, sent an "URGENT: Medical Device Product Removal" letter dated October 6, 2010, to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1) Screen their inventory immediately for any of the part numbers listed in the letter. Do not include any product received after October 7, 2010. 2) If they have any of the products listed, contact Anspach Customer Support at (800) 327-6887 to arrange for their immediate return and replacement. 3) Complete and return the attached Customer Reply Form indicating how many of each part number is being returned and confirming their receipt of this letter, and fax to 1-800-327-6661. If they have distributed any of the products to other services or facilities, please forward this information as appropriate. Should you have any queries, please do not hesitate to contact Anspach Customer Support at (800) 327-6887.
Quantity
271 cutting burs affected
Distribution
Worldwide distribution: USA including states of: AZ, CA, FL, GA, IL, IN, LA, MA, and MI; and countries of: Australia, Belgium, England, France, Italy, Japan, South Africa, Spain and Sweden.
Lot, serial or model codes
Lot numbers: C523029666, D063031864, D113032937, D113032974, D163033908, D253036524, and D353039102.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTT
Firm FEI
1045834
Firm city
Palm Beach Gardens
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register